Investigation of retention on bare silica using reversed-phase mobile phases at elevated temperatures

被引:34
作者
Bidlingmeyer, BA [1 ]
Henderson, J [1 ]
机构
[1] Agilent Technol, Wilmington, DE 19808 USA
关键词
temperature effects; Van't Hoff plot; silica; retention behavior;
D O I
10.1016/j.chroma.2004.08.113
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
The use of unbonded silica as a stationary phase in reversed-phase HPLC is described as a useful alternative to bonded phase columns for polar, lipophilic amines. Using four lipophilic amines, the role of temperature is shown to favorably impact both efficiency and selectivity, which is not universally seen when using bonded phases. As temperature is raised, retention drops on the silica column. The temperature behavior appears to support the hypothesis that retention is dependant upon electrostatic and adsorptive forces. (C) 2004 Elsevier B.V. All rights reserved.
引用
收藏
页码:187 / 193
页数:7
相关论文
共 36 条
[1]   IMPROVEMENTS IN LIQUID-CHROMATOGRAPHY COLUMN LIFE AND METHOD FLEXIBILITY BY SATURATING THE MOBILE PHASE WITH SILICA [J].
ATWOOD, JG ;
SCHMIDT, GJ ;
SLAVIN, W .
JOURNAL OF CHROMATOGRAPHY, 1979, 171 (APR) :109-115
[2]   PREPARATION OF VARIOUS BONDED PHASES FOR HPLC USING MONOCHLOROSILANES [J].
BERENDSEN, GE ;
PIKAART, KA ;
GALAN, LD .
JOURNAL OF LIQUID CHROMATOGRAPHY, 1980, 3 (10) :1437-1464
[3]   Trends in reversed-phase HPLC column practices: 1997 [J].
Bidlingmeyer, BA .
JOURNAL OF CHROMATOGRAPHIC SCIENCE, 1997, 35 (08) :392-400
[4]   SEPARATION OF ORGANIC AMINE COMPOUNDS ON SILICA-GEL WITH REVERSED-PHASE ELUENTS [J].
BIDLINGMEYER, BA ;
DELRIOS, JK ;
KORPI, J .
ANALYTICAL CHEMISTRY, 1982, 54 (03) :442-447
[5]  
BIDLINGMEYER BA, 1996, J CHROMATOGR, V34, P307
[6]  
BIDLINGMEYER BA, 1969, SEP SCI, V4, P439
[7]  
BIJ K, 1979, J CHROMATOGR, V185, P129
[8]  
BOUNINE JP, 1984, J CHROMATOGR, V298, P1, DOI 10.1016/S0021-9673(01)92691-3
[9]   Effect of phase ratio on van't Hoff analysis in reversed-phase liquid chromatography, and phase-ratio-independent estimation of transfer enthalpy [J].
Chester, TL ;
Coym, JW .
JOURNAL OF CHROMATOGRAPHY A, 2003, 1003 (1-2) :101-111
[10]   Effect of buffers on silica-based column stability in reversed-phase high-performance liquid chromatography [J].
Claessens, HA ;
vanStraten, MA ;
Kirkland, JJ .
JOURNAL OF CHROMATOGRAPHY A, 1996, 728 (1-2) :259-270