Gradient-echo line scan imaging using 2D-Selective RF excitation

被引:7
作者
Finsterbusch, J [1 ]
Frahm, J [1 ]
机构
[1] Max Planck Inst Biophys Chem, Biomed NMR Forsch GmbH, D-37070 Gottingen, Germany
关键词
magnetic resonance imaging; line scan imaging; 2D-selective RF excitation; inner-volume imaging; fMRI;
D O I
10.1006/jmre.2000.2159
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
A gradient-echo line scan imaging technique was developed which employs two-dimensional spatially selective radiofrequency (2DRF) pulses for consecutively exciting individual columns of transverse magnetization, i.e., image lines. Although a variety of trajectories are possible for 2DRF excitation, the current implementation involved a blipped-planar trajectory in conjunction with additional saturation RF pulses to suppress side excitations above and below the desired image section, i.e., along the blip direction of the 2DRF pulse. Human brain imaging at 2.0 T (Siemens Vision, Erlangen, Germany) resulted in measuring times of 5.2 s for a 5-mm section at 1.0 x 1.0 mm in-plane resolution. Functional neuroimaging of the motor cortex at 1.2 s temporal resolution and 0.78 x 1.56 mm in-plane resolution exploited the capability of imaging inner volumes (here a 25-mm strip) without signal aliasing. (C) 2000 Academic Press.