Plasma-deposited tetraglyme surfaces greatly reduce total blood protein adsorption, contact activation, platelet adhesion, platelet procoagulant activity, and in vitro thrombus deposition

被引:65
作者
Cao, Lan
Chang, Mark
Lee, Chi-Ying
Castner, David G.
Sukavaneshvar, Sivaprasad
Ratner, Buddy D.
Horbett, Thomas A. [1 ]
机构
[1] Univ Washington, Dept Chem Engn, Seattle, WA 98195 USA
[2] Univ Washington, Dept Bioengn, Seattle, WA 98195 USA
[3] Utah Artificial Heart Inst, Salt Lake City, UT 84103 USA
关键词
tetraglyme; protein adsorption; platelet adhesion; platelet activation; plasma deposition;
D O I
10.1002/jbm.a.31091
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The ability of tetraethylene glycol dimethyl ether (tetraglyme) plasma deposited coatings exhibiting ultralow fibrinogen adsorption to reduce blood activation was studied with six in vitro methods, namely fibrinogen and von Willebrand's factor adsorption, total protein adsorption, clotting time in recalcified plasma, platelet adhesion and procoagulant activity, and whole blood thrombosis in a disturbed flow catheter model. Surface plasmon resonance results showed that tetraglyme surfaces strongly resisted the adsorption of all proteins from human plasma. The clotting time in the presence of tetraglyme surfaces was lengthened compared with controls, indicating a lower activation of the intrinsic coagulation cascade. Platelet adhesion and thrombin generation by adherent platelets were greatly reduced on tetraglyme-coated materials, compared with uncoated and Biospan (R)-coated glass slides. in the in vitro disturbed blood flow model, tetraglyme plasma coated catheters had 50% less thrombus than did the uncoated catheters. Tetraglyme-coated materials thus had greatly reduced blood interactions as measured with all six methods. The improved blood compatibility of plasma-deposited tetraglyme is thus not only due to their reduced platelet adhesion and activation, but also to a generalized reduction in blood interactions. (c) 2007 Wiley Periodicals, Inc. J Biomed Mater Res 81A: 827-837, 2007.
引用
收藏
页码:827 / 837
页数:11
相关论文
共 47 条
  • [1] [Anonymous], BIOMATERIALS SCI INT
  • [2] ANTONSEN KP, 1992, TOPICS APPL CHEM
  • [3] Colloid probe AFM investigation of interactions between fibrinogen and PEG-like plasma polymer surfaces
    Bremmell, KE
    Kingshott, P
    Ademovic, Z
    Winther-Jensen, B
    Griesser, HJ
    [J]. LANGMUIR, 2006, 22 (01) : 313 - 318
  • [4] CAO L, IN PRESS J BIOMED MA
  • [5] The effect of adsorbed fibrinogen, fibronectin, von Willebrand factor and vitronectin on the procoagulant state of adherent platelets
    Grunkemeier, JM
    Tsai, WB
    McFarland, CD
    Horbett, TA
    [J]. BIOMATERIALS, 2000, 21 (22) : 2243 - 2252
  • [6] Grunkemeier JM, 1998, J BIOMED MATER RES, V41, P657, DOI 10.1002/(SICI)1097-4636(19980915)41:4<657::AID-JBM18>3.0.CO
  • [7] 2-B
  • [8] Grunkemeier JM, 2000, J BIOMED MATER RES, V51, P669, DOI 10.1002/1097-4636(20000915)51:4<669::AID-JBM15>3.0.CO
  • [9] 2-#
  • [10] HANSON SR, 1980, J LAB CLIN MED, V95, P289