Comparison of AutoSet™ and polysomnography for the detection of apnea-hypopnea events

被引:7
作者
Bagnato, MC
Nery, LE
Moura, SMT
Bittencourt, LRA
Tufik, S
机构
[1] Univ Fed Sao Paulo, Dept Psicobiol, BR-04024002 Sao Paulo, Brazil
[2] Univ Fed Sao Paulo, Disciplina Pneumol, BR-04024002 Sao Paulo, Brazil
关键词
obstructive steep apnea; polysomnography; computerized diagnostic system;
D O I
10.1590/S0100-879X2000000500005
中图分类号
Q [生物科学];
学科分类号
07 ; 0710 ; 09 ;
摘要
The use of the flow vs time relationship obtained with the nasal prongs of the AutoSet(TM) (AS) system (diagnosis mode) has been proposed to detect apneas and hppopneas in patients with reasonable nasal patency. Our aim was to compare the accuracy of AS to that of a computerized polysomnoeraphic (PSG) system. The study was conducted on 56 individuals (45 men) with clinical characteristics of obstructive sleep apnea (OSA), Their mean (+/- SD) age was 44.6 +/- 12 years and their body mass index was 31.3 +/- 7 kg/m(2). Data were submitted to parametric analysis to determine the agreement between methods and the intraclass correlation coefficient was calculated. The Student t-test and Bland and Altman plots were also used. Twelve patients had an apnea-hypopnea index (AHI) <10 in bed and 20 had values >40. The mean (+/- SD) AHI(PSG) index of 37.6 (28.8) was significantly lower (P = 0.0003) than AHI(AS) (41.8 (25.3)), but there was a high intraclass correlation coefficient (0.93), with 0.016 variance. For a threshold of AHI of 20, AS showed 73.0% accuracy, 97% sensitivity and 60% specificity, with positive and negative predictive values of 78% and 93%, respectively. Sensitivity, specificity and negative predictive values increased in parallel to the increase in AHI threshold for detecting OSA. However, when the differences of AHI(PSG-AS) were plotted against their means, the limits of agreement between the methods (95% of the differences) were +13 and -22, showing the discrepancy between the AHI values obtained with PSG and AS. Finally, cubic regression analysis was used to better predict the result of AHI(PSG) as a function of the method proposed, i.e., AHI(AS). We conclude that, despite these differences, AHI measured by AutoSet(TM) can be useful for the assessment of patients with high pre-test clinical probability of OSA, for whom standard PSG is not possible as an initial step in diagnosis.
引用
收藏
页码:515 / 519
页数:5
相关论文
共 17 条
[1]  
Bagnato M. C., 1996, Journal of Sleep Research, V5, P11
[2]  
BERTHONJONES M, 1993, SLEEP, V16, P120
[3]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[4]   FACTORS IMPAIRING DAYTIME PERFORMANCE IN PATIENTS WITH SLEEP-APNEA HYPOPNEA SYNDROME [J].
CHESHIRE, K ;
ENGLEMAN, H ;
DEARY, I ;
SHAPIRO, C ;
DOUGLAS, NJ .
ARCHIVES OF INTERNAL MEDICINE, 1992, 152 (03) :538-541
[5]   EFFECT OF CONTINUOUS POSITIVE AIRWAY PRESSURE TREATMENT ON DAYTIME FUNCTION IN SLEEP APNOEA/HYPOPNOEA SYNDROME [J].
ENGLEMAN, HM ;
MARTIN, SE ;
DEARY, IJ ;
DOUGLAS, NJ .
LANCET, 1994, 343 (8897) :572-575
[6]   AUTOMOBILE ACCIDENTS INVOLVING PATIENTS WITH OBSTRUCTIVE SLEEP-APNEA [J].
FINDLEY, LJ ;
UNVERZAGT, ME ;
SURATT, PM .
AMERICAN REVIEW OF RESPIRATORY DISEASE, 1988, 138 (02) :337-340
[7]   A laboratory validation study of the diagnostic mode of the Autoset(TM) system for sleep-related respiratory disorders [J].
Fleury, B ;
Rakotonanahary, D ;
HausserHauw, C ;
Lebeau, B ;
Guilleminault, C .
SLEEP, 1996, 19 (06) :502-505
[8]  
Gugger M, 1997, EUR RESPIR J, V10, P587
[9]   A CAUSE OF EXCESSIVE DAYTIME SLEEPINESS - THE UPPER AIRWAY-RESISTANCE SYNDROME [J].
GUILLEMINAULT, C ;
STOOHS, R ;
CLERK, A ;
CETEL, M ;
MAISTROS, P .
CHEST, 1993, 104 (03) :781-787
[10]   EVALUATION OF SLEEP-DISORDERED BREATHING - IS POLYSOMNOGRAPHY NECESSARY [J].
HAPONIK, EF ;
SMITH, PL ;
MEYERS, DA ;
BLEECKER, ER .
AMERICAN JOURNAL OF MEDICINE, 1984, 77 (04) :671-677