The effect of implant size and device keel on vertebral compression properties in lumbar total disc replacement

被引:39
作者
Auerbach, Joshua D. [1 ,2 ]
Ballester, Carrie M. [2 ]
Hammond, Frank [2 ]
Carine, Ehren T. [2 ]
Balderston, Richard A. [3 ]
Elliott, Dawn M. [2 ]
机构
[1] Albert Einstein Coll Med, Bronx Lebanon Hosp Ctr, Dept Orthopaed, Bronx, NY 10457 USA
[2] Univ Penn, McKay Orthopaed Res Lab, Dept Orthopaed Surg, Philadelphia, PA 19104 USA
[3] Penn Hosp, Philadelphia, PA 19104 USA
关键词
Lumbar total disc arthroplasty; End plate; Compression; Mechanics; Keel; ADJACENT MOTION SEGMENTS; FOLLOW-UP PERIOD; ARTIFICIAL DISC; END-PLATE; CIRCUMFERENTIAL FUSION; INTRADISCAL PRESSURE; PRODISC-II; ARTHROPLASTY; PROSTHESIS; STRENGTH;
D O I
10.1016/j.spinee.2010.01.008
中图分类号
R74 [神经病学与精神病学];
学科分类号
100204 [神经病学];
摘要
BACKGROUND CONTEXT: Vertebral end plate support is necessary for successful lumbar total disc replacement (TDR) surgery. Failure to achieve anterior column support as a result of lumbar TDR device undersizing could lead to implant subsidence and fracture. PURPOSE: The purpose of the study was to examine the compressive biomechanical behavior of the vertebral end plate with varying sizes of disc replacement implants. STUDY DESIGN: The study design comprises a biomechanical investigation using a human cadaveric lumbar spine model. METHODS: Fifty-six vertebrae with intact posterior elements were prepared from 13 fresh frozen lumbar spines. Peripheral quantitative computed tomography was performed to assess regional bone density. Vertebrae were potted and subjected to nondestructive compression testing with a !;mall, medium, and large custom-made implants with the footplate geometry of the ProDisc-L TDR (Synthes Spine, West Chester, PA, USA) system and having no keel. Failure testing was performed using the ProDisc-L implant with an intact keel. Pressure sensor film was used to assess contact pressure and distribution. RESULTS: There was a linear correlation between percent coverage of the end plate and implant-end plate stiffness (p=.0001) and an inverse correlation with displacement (p=.01). The difference in implant-end plate stiffness between small-medium, medium-large, and small-large implants was 10.5% (p=.03), 10.2% (p=.02), and 19.6% (p<.0001), respectively. Failure analysis revealed similar trends for implant sizing, but only bone density was found to significantly correlate with failure properties (r=0.76, p<.0001). There was a significant reduction in implant-end plate stiffness of 18% when the keel was intact compared to without the keel (range 6-27%, p=.0008). Pressure film analysis revealed that the implant was loaded peripherally and did not have central contact during nondestructive loading. There was a trend toward greater contact pressure with the small implant when compared with the medium implant (p=.06) and the large implant (p=.06). CONCLUSIONS: Although larger implants reduce end plate displacement, increase apparent implant-end plate stiffness, increase the implant-end plate contact area, and decrease the peak contact pressures, low bone density reduces failure properties. The keel introduces a reduction in stiffness to the implant-end plate interface, the clinical significance of which is currently unknown. (C) 2010 Elsevier Inc. All rights reserved.
引用
收藏
页码:333 / 340
页数:8
相关论文
共 46 条
[1]
AUERBACH JD, 2007, P 34 INT SOC STUD LU, P151
[2]
Evaluation of spinal kinematics following lumbar total disc replacement and circumferential fusion using in vivo fluoroscopy [J].
Auerbach, Joshua D. ;
Wills, Brian P. D. ;
McIntosh, Theresa C. ;
Balderston, Richard A. .
SPINE, 2007, 32 (05) :527-536
[3]
Caspi I, 2003, ISRAEL MED ASSOC J, V5, P9
[4]
Results of disc prosthesis after a minimum follow-up period of 2 years [J].
Cinotti, G ;
David, T ;
Postacchini, F .
SPINE, 1996, 21 (08) :995-1000
[5]
Biomechanical evaluation of total disc replacement arthroplasty:: An in vitro human cadaveric model [J].
Cunningham, BW ;
Gordon, JD ;
Dmitriev, AE ;
Hu, NB ;
McAfee, PC .
SPINE, 2003, 28 (20) :S110-S117
[6]
Long-term results of one-level lumbar arthroplasty - Minimum 10-year follow-up of the CHARITE artificial disc in 106 patients [J].
David, Thierry .
SPINE, 2007, 32 (06) :661-666
[7]
Clinical results of ProDisc-II lumbar total disc replacement: Report from the United States clinical trial [J].
Delamarter, RB ;
Bae, HW ;
Pradhan, BB .
ORTHOPEDIC CLINICS OF NORTH AMERICA, 2005, 36 (03) :301-+
[8]
Adjacent level intradiscal pressure and segmental kinematics following a cervical total disc arthroplasty -: An In Vitro human cadaveric model [J].
Dmitriev, AE ;
Cunningham, BW ;
Hu, NB ;
Sell, G ;
Vigna, F ;
McAfee, PC .
SPINE, 2005, 30 (10) :1165-1172
[9]
Adjacent segment degeneration in the lumbar spine [J].
Ghiselli, G ;
Wang, JC ;
Bhatia, NN ;
Hsu, WK ;
Dawson, EG .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2004, 86A (07) :1497-1503
[10]
The fate of the adjacent motion segments after lumbar fusion [J].
Gillet, P .
JOURNAL OF SPINAL DISORDERS & TECHNIQUES, 2003, 16 (04) :338-345