The effect of a model-based predictive display on the control of end-tidal sevoflurane concentrations during low-flow anesthesia

被引:30
作者
Kennedy, RR
French, RA
Gilles, S
机构
[1] Christchurch Hosp, Dept Anaesthesia, Christchurch, New Zealand
[2] Christchurch Sch Med & Hlth Sci, Christchurch, New Zealand
关键词
D O I
10.1213/01.ANE.0000131507.65132.7F
中图分类号
R614 [麻醉学];
学科分类号
100217 [麻醉学];
摘要
We have shown that a multicompartment model accurately predicts end-tidal (ET) sevoflurane (sevo) and isoflurane concentrations. The model has been adapted to use real-time fresh gas flow and vaporizer settings to display a 10-min prediction of ET sevo concentrations. In this study, we evaluated the effect of the predictive display on the speed and accuracy of changes in ET sevo by the anesthesiologist. Fifteen patients were studied in whom sevo-based anesthesia was expected to last more than 2 h. Four step changes of target ET concentration (+0.5, +1.0, -1.0, and -0.5 vol%) were made either unaided or with the prediction display. Fresh gas flow was 1 L/min. Response time, maximum overshoot, and stability in the 5 min after the target was achieved were compared by using two-tailed paired Student's t-tests. Changes were made on average 1.5-2.3 times faster with the predictive display than without it. These differences were statistically significant (P < 0.05) for the +0.5, +1.0, and -0.5 vol% step changes but not for the -1.0 vol% change. There were no differences in the degree of overshoot or stability. These differences are comparable to those seen with an automatic feedback control system. This system may simplify the administration of volatile anesthesia and the use of low-flow anesthesia.
引用
收藏
页码:1159 / 1163
页数:5
相关论文
共 20 条
[1]
A manual slide rule for target-controlled infusion of propofol:: Development and evaluation [J].
Bruhn, J ;
Bouillon, TW ;
Röpcke, H ;
Hoeft, A .
ANESTHESIA AND ANALGESIA, 2003, 96 (01) :142-147
[2]
THE UPTAKE OF ENFLURANE DURING ANESTHESIA [J].
DEVCHAND, D ;
CHAKRABARTI, MK ;
LOCKWOOD, GG .
ANAESTHESIA, 1995, 50 (06) :491-495
[3]
Target-controlled infusions for intravenous anesthetics - Surfing USA not! [J].
Egan, TD ;
Shafer, SL .
ANESTHESIOLOGY, 2003, 99 (05) :1039-1041
[4]
GAGE JS, 1992, M D COMPUT, V9, P54
[5]
HEFFERNAN PB, 1982, ANAESTHESIA, V37, P9, DOI 10.1111/j.1365-2044.1982.tb00986.x
[6]
Kennedy Ross, 2003, N Z Med J, V116, pU438
[7]
Predictive accuracy of a model of volatile anesthetic uptake [J].
Kennedy, RR ;
French, RA ;
Spencer, C .
ANESTHESIA AND ANALGESIA, 2002, 95 (06) :1616-1621
[8]
The effect of cardiac output changes on end-expired volatile anaesthetic concentrations - a theoretical study [J].
Kennedy, RR ;
Baker, AB .
ANAESTHESIA, 2001, 56 (11) :1034-1040
[9]
SOLUBILITY CHARACTERISTICS OF THE IDEAL AGENT FOR MEASUREMENT OF CARDIAC-OUTPUT BY SOLUBLE GAS UPTAKE METHODS [J].
KENNEDY, RR ;
BAKER, AB .
BRITISH JOURNAL OF ANAESTHESIA, 1993, 71 (03) :398-402
[10]
A SYSTEM MODEL FOR CLOSED-CIRCUIT INHALATION ANESTHESIA .1. COMPUTER STUDY [J].
LEROU, JGC ;
DIRKSEN, R ;
KOLMER, HHB ;
BOOIJ, LHDJ .
ANESTHESIOLOGY, 1991, 75 (02) :345-355