Comparative assessment of chemical and γ-irradiation procedures for implantable glucose enzyme electrodes

被引:6
作者
Ahmed, S [1 ]
Rigby, GP [1 ]
Crump, P [1 ]
Vadgama, PM [1 ]
机构
[1] Hope Hosp, Clin Biochem Sect, Salford M6 8HD, Lancs, England
关键词
needle glucose sensors; sterilisation; irradiation;
D O I
10.1016/S0956-5663(00)00057-9
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
Chemical and gamma-irradiation sterilisation were examined in this study for implantable needle electrodes. Exposure to isopropyl alcohol (IPA) led to response elevation, but time-dependent exposures up to 30 min variously to chlorhexidine, H2O2, HCl, HCl/IPA, and alcoholic iodine/potassium iodide, all caused substantial time-dependent response degradation. Sterility was not assessed for such electrodes. High dose (30 KGy) gamma-irradiation also compromised sensor response, with exposed electrodes exhibiting similar to 30% reduction in response, comparable in magnitude to 10 min exposure to chemical sterilisation. However, a lower dose (25 KGy) gamma-irradiation was well tolerated and subsequent microbiological testing (following contamination with Streptococcus epidermidis and Staphylococus aureus) proved device sterility with no culture growth following 7 days incubation at 37 degrees C in brain heart infusion medium. (C) 2000 Elsevier Science S.A. All rights reserved.
引用
收藏
页码:159 / 165
页数:7
相关论文
共 17 条
[1]   DISINFECTION OF HEAT-SENSITIVE MATERIAL BY LOW-TEMPERATURE STEAM AND FORMALDEHYDE [J].
ALDER, VG ;
BROWN, AM ;
GILLESPIE, WA .
JOURNAL OF CLINICAL PATHOLOGY, 1966, 19 (01) :83-+
[2]  
BINDRA SD, 1991, ANAL CHEM, V63, P1693
[3]   NEEDLE ENZYME ELECTRODES FOR BIOLOGICAL STUDIES [J].
CHURCHOUSE, SJ ;
BATTERSBY, CM ;
MULLEN, WH ;
VADGAMA, PM .
BIOSENSORS, 1986, 2 (06) :325-342
[4]  
KLEEPE K, 1966, BIOCHEMISTRY-US, V5, P139
[5]  
MALITESTA C, 1990, ANAL CHEM, V62, P24
[6]   EXTRACORPOREAL DETERMINATION OF GLUCOSE, LACTATE AND POTASSIUM WITH ELECTROCHEMICAL BIOSENSORS [J].
MASCINI, M ;
PALLESCHI, G ;
MOSCONE, D ;
BERNARDI, L .
JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 1989, 7 (12) :1377-1383
[7]   EFFECT OF PERIODATE OXIDATION ON STRUCTURE AND PROPERTIES OF GLUCOSE OXIDASE [J].
NAKAMURA, S ;
HAYASHI, S ;
KOGA, K .
BIOCHIMICA ET BIOPHYSICA ACTA, 1976, 445 (02) :294-308
[8]  
ODOHERTY J, 1997, J MED DEV TECH, V8, P16
[9]  
ODONNELL JH, 1989, ACS SYM SER, V381, P1
[10]  
Richards S, 1996, Med Device Technol, V7, P22