Polymeric micro-cantilever array for auditory front-end processing

被引:42
作者
Xu, T [1 ]
Bachman, M
Zeng, FG
Li, GP
机构
[1] Univ Calif Irvine, Dept Biomed Engn, Irvine, CA 92697 USA
[2] Univ Calif Irvine, Dept Elect & Comp Engn, Irvine, CA 92697 USA
[3] Univ Calif Irvine, Dept Otolaryngol, Irvine, CA 92697 USA
关键词
cantilever array; polymeric cantilever; acoustic sensor; auditory prostheses; mechanical filter; cochlear implant; hearing aid;
D O I
10.1016/j.sna.2003.11.035
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
A polymeric micro-cantilever array has been developed that mimics the biological front-end processing in the mammalian cochlea, and is intended for use in applications of auditory prostheses. Made of optical epoxy polymer molded over a silicon substrate, the micro-cantilevers have similar mechanical performance as the basilar membranes in the mammalian ears. The polymeric cantilevers, which are transparent, are used as optical waveguides to guide and modulate a light beam, which is initiated from a light emitting diode (LED) and collected by a photo diode, to produce a signal suitable for human hearing when the sound wave excites them. The polymeric cantilevers have Q(10) values of 9.38, 10.11, 11.56, and 14.01 for resonant frequencies at 286, 720, 2868, and 6948 Hz, respectively. These values are similar to those obtained by direct measurement of the basilar membrane. Furthermore, they have a linear dynamic range of more than 80 dB sound pressure level (SPL) with less than 15% total harmonic distortion (THD). This polymeric micro-cantilever array has low power consumption, short processing time, high sensitivity, high frequency resolution, small size, is insensitive to electromagnetic interference, and is suitable for a totally implantable device in the human ear. (C) 2003 Elsevier B.V. All rights reserved.
引用
收藏
页码:176 / 182
页数:7
相关论文
共 11 条
[1]   Hearing with bionic ears [J].
Ay, SU ;
Zeng, FG ;
Sheu, BJ .
IEEE CIRCUITS & DEVICES, 1997, 13 (03) :18-23
[2]   AUDITORY PROSTHESES - IMPLANTABLE AND VIBROTACTILE DEVICES [J].
DORMER, KJ ;
PHILLIPS, MA .
IEEE ENGINEERING IN MEDICINE AND BIOLOGY MAGAZINE, 1987, 6 (02) :36-41
[3]   Fish-bone structured acoustic sensor toward silicon cochlear systems [J].
Harada, M ;
Ikeuchi, N ;
Fukui, S ;
Ando, S .
MICROMACHINED DEVICES AND COMPONENTS IV, 1998, 3514 :266-275
[4]  
Harada Muneo, 1999, T IEE JAPAN D, V119, P125
[5]  
LOIZOU PC, 1997, P MIDW S CIRC SYST M, P200
[6]  
McDermott H, 1998, IEEE Trans Rehabil Eng, V6, P53, DOI 10.1109/86.662620
[7]  
Muller M, 1998, P SOC PHOTO-OPT INS, V3514, P322, DOI 10.1117/12.323905
[8]   Mechanics of the mammalian cochlea [J].
Robles, L ;
Ruggero, MA .
PHYSIOLOGICAL REVIEWS, 2001, 81 (03) :1305-1352
[9]   Design and fabrication of a MEMS filter bank for hearing aids applications [J].
Shen, SH ;
Young, ST ;
Fang, WL .
2ND ANNUAL INTERNATIONAL IEEE-EMBS SPECIAL TOPIC CONFERENCE ON MICROTECHNOLOGIES IN MEDICINE & BIOLOGY, PROCEEDINGS, 2002, :352-355
[10]  
Wise KD, 2002, INTERNATIONAL ELECTRON DEVICES 2002 MEETING, TECHNICAL DIGEST, P499, DOI 10.1109/IEDM.2002.1175889