New controlled environment vitrification system for cryo-transmission electron microscopy:: design and application to surfactant solutions

被引:19
作者
Egelhaaf, SU
Schurtenberger, P
Müller, M
机构
[1] ETH Zurich, Labor Elektronenmikroskopie 1, CH-8092 Zurich, Switzerland
[2] ETH Zurich, Inst Polymere, CH-8092 Zurich, Switzerland
来源
JOURNAL OF MICROSCOPY-OXFORD | 2000年 / 200卷 / 02期
关键词
bile salt; controlled environment vitrification system (CEVS); cryo-transmission electron microscopy; lecithin; light scattering; micelles; sample preparation; surfactants; transfer system;
D O I
10.1046/j.1365-2818.2000.00747.x
中图分类号
TH742 [显微镜];
学科分类号
摘要
A newly designed controlled environment vitrification system for cryo-transmission electron microscopy of vitrified samples (cryo TEM) is presented. It significantly reduces the lag-time from blotting to Vitrification by means of an optimized blotting procedure. Furthermore, a modified transfer system allows rapid transfer of the vitrified sample into the microscope and reduces the equilibration time. The practicable throughput of samples is thereby markedly increased. This new equipment is used to determine the overall size of individual polymer-like mixed lecithin-bile salt micelles, which very critically depends on the composition of the sample. The quantitative agreement with the results from independent light scattering experiments an the same sample supports the validity and performance of the presented sample preparation and handling procedures. In addition, these experiments demonstrate that cryo TEM experiments not only provide direct structural information, but furthermore permit a quantitative determination of micellar properties.
引用
收藏
页码:128 / 139
页数:12
相关论文
共 55 条
[1]   Cryotransmission electron microscopy of thin vitrified samples [J].
Almgren, M ;
Edwards, K ;
Gustafsson, J .
CURRENT OPINION IN COLLOID & INTERFACE SCIENCE, 1996, 1 (02) :270-278
[2]   VITRIFICATION OF AQUEOUS SUSPENSIONS FROM A CONTROLLED ENVIRONMENT FOR ELECTRON-MICROSCOPY - AN IMPROVED PLUNGE COOLING DEVICE [J].
BATTERSBY, BJ ;
SHARP, JCW ;
WEBB, RI ;
BARNES, GT .
JOURNAL OF MICROSCOPY-OXFORD, 1994, 176 :110-120
[3]   CONTROLLED ENVIRONMENT VITRIFICATION SYSTEM - AN IMPROVED SAMPLE PREPARATION TECHNIQUE [J].
BELLARE, JR ;
DAVIS, HT ;
SCRIVEN, LE ;
TALMON, Y .
JOURNAL OF ELECTRON MICROSCOPY TECHNIQUE, 1988, 10 (01) :87-111
[4]  
BURCHARD W, 1983, ADV POLYM SCI, V48, P1
[5]   FREEZE-FRACTURE-REPLICATION USING THE CONTROLLED ENVIRONMENT VITRIFICATION SYSTEM (CEVS) [J].
BURNS, JL ;
TALMON, Y .
JOURNAL OF ELECTRON MICROSCOPY TECHNIQUE, 1988, 10 (01) :113-114
[6]   STATICS AND DYNAMICS OF WORM-LIKE SURFACTANT MICELLES [J].
CATES, ME ;
CANDAU, SJ .
JOURNAL OF PHYSICS-CONDENSED MATTER, 1990, 2 (33) :6869-6892
[7]  
CHAMBERLIN RA, 1991, THESIS MIT CAMBRIDGE
[8]  
COHEN DE, 1991, FALK S, V58, P147
[9]   PARTICLE-SURFACE INTERACTION IN THIN VITRIFIED FILMS FOR CRYOELECTRON MICROSCOPY [J].
CYRKLAFF, M ;
ROOS, N ;
GROSS, H ;
DUBOCHET, J .
JOURNAL OF MICROSCOPY-OXFORD, 1994, 175 :135-142
[10]  
de Gennes P.G., 1979, SCALING CONCEPTS POL