Magnetoencephalographic validation parameters for clinical evaluation of interictal epileptic activity

被引:30
作者
Bowyer, SM
Mason, K
Tepley, N
Smith, B
Barkley, GL
机构
[1] Henry Ford Hlth Syst, Dept Neurol, Detroit, MI USA
[2] Oakland Univ, Dept Phys, Rochester, MI USA
[3] Case Western Reserve Univ, Dept Neurol, Cleveland, OH 44106 USA
关键词
magnetoencephalography; epilepsy; equivalent current dipole; confidence volume;
D O I
10.1097/00004691-200304000-00001
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
The authors demonstrate that the confidence volume (the spatial volume that encompasses the 95% probability of source localization) of the single equivalent current dipole is helpful invalidating magnetoencephalographic epileptic spike mapping. Such mapping involves distinguishing spikes from other neuronal events. The usual criteria for validating dipole fit reliability involve four parameters-correlation coefficient (R greater than or equal to 0.98), goodness of fit (greater than or equal to 0.95), root mean square magnetic field value (> 400 fT), and dipole moment (Q value > 200 nAm)-but other parameters (direction of dipole moment, location of dipole, and confidence volume) can be considered. In 21 patients with epilepsy, the average correlation coefficient for 608 epileptic spikes was 0.99; average goodness of fit, 0.98; average root mean square, 1,198 fT; and the average Q value, 370 nAm. The mean average confidence volume was 0.30 +/- 0.27 cm(3). Correlation coefficient values for quiet brain activity were less than 0.90; goodness of fit values, less than 0.85; and confidence volumes were large (> 5 cm(3)); and for noise runs (no subject) they were even larger (> 100 cm(3)), although correlation coefficient values were more than 0.80 and goodness of fit values were more than 0.85. Confidence volumes for noise data are large-for background brain activity even larger-but confidence volumes for epileptic spikes are small. Confidence volume, in conjunction with other parameters, may be a robust parameter for spike selection.
引用
收藏
页码:87 / 93
页数:7
相关论文
共 12 条
  • [1] BOWYER SM, 2001, BIOMAG 2000 P 12 INT, P484
  • [2] MATHEMATICAL DIPOLES ARE ADEQUATE TO DESCRIBE REALISTIC GENERATORS OF HUMAN-BRAIN ACTIVITY
    DEMUNCK, JC
    VANDIJK, BW
    SPEKREIJSE, H
    [J]. IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1988, 35 (11) : 960 - 966
  • [3] Ebersole JS, 1997, EPILEPSIA S4, V38, P1
  • [4] Ebersole JS, 1997, COMPREHENSIVE EPILEP, P919
  • [5] PRESURGICAL LOCALIZATION OF FUNCTIONAL CORTEX USING MAGNETIC SOURCE IMAGING
    GALLEN, CC
    SCHWARTZ, BJ
    BUCHOLZ, RD
    MALIK, G
    BARKLEY, GL
    SMITH, J
    TUNG, H
    COPELAND, B
    BRUNO, L
    ASSAM, S
    HIRSCHKOFF, E
    BLOOM, F
    [J]. JOURNAL OF NEUROSURGERY, 1995, 82 (06) : 988 - 994
  • [6] MAGNETOENCEPHALOGRAPHY - THEORY, INSTRUMENTATION, AND APPLICATIONS TO NONINVASIVE STUDIES OF THE WORKING HUMAN BRAIN
    HAMALAINEN, M
    HARI, R
    ILMONIEMI, RJ
    KNUUTILA, J
    LOUNASMAA, OV
    [J]. REVIEWS OF MODERN PHYSICS, 1993, 65 (02) : 413 - 497
  • [7] KAUKORANTA E, 1986, EXP BRAIN RES, V63, P60
  • [8] Postoperative multichannel magnetoencephalography in patients with recurrent seizures after epilepsy surgery
    Kirchberger, K
    Hummel, C
    Stefan, H
    [J]. ACTA NEUROLOGICA SCANDINAVICA, 1998, 98 (01): : 1 - 7
  • [9] Lypchuk T., 1990, 4D NEUROIMAGING SOFT, P1
  • [10] The correlation coefficient and the goodness of fit in source localization of noise recorded by magnetoencephalography
    Siedenberg, R
    Goodin, DS
    Aminoff, MJ
    Rowley, HA
    Roberts, TPI
    [J]. BRAIN TOPOGRAPHY, 1996, 9 (02) : 95 - 100