Reproducibility of Lung and Lobar Volume Measurements Using Computed Tomography

被引:47
作者
Brown, Matthew S. [1 ]
Kim, Hyun J. [1 ]
Abtin, Fereidoun [1 ]
Da Costa, Irene [1 ]
Pais, Richard [1 ]
Ahmad, Shama [1 ]
Angel, Erin [1 ]
Ni, Chiayi [1 ]
Kleerup, Eric C. [2 ]
Gjertson, David W.
McNitt-Gray, Michael F. [1 ]
Goldin, Jonathan G. [1 ]
机构
[1] Univ Calif Los Angeles, Dept Radiol Sci, Thorac Imaging Res Grp, Los Angeles, CA 90024 USA
[2] Univ Calif Los Angeles, Div Pulm & Crit Care Med, Los Angeles, CA 90024 USA
关键词
Body plethysmography; emphysema; image analysis; PULMONARY-EMPHYSEMA; CT; SEGMENTATION; QUANTIFICATION; GAS;
D O I
10.1016/j.acra.2009.10.005
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
100231 [临床病理学]; 100902 [航空航天医学];
摘要
Rationale and Objectives: Lung and lobar volume measurements from computed tomographic (CT) imaging are being used in clinical trials to assess new minimally invasive emphysema treatments aiming to reduce lung volumes. Establishing the reproducibility of lung volume measurements is important if they are to be accepted as treatment planning and outcome variables. The aims of this study were to (1) investigate the correlation between lung volumes assessed on CT imaging and on pulmonary function testing (PFT), (2) compare the two methods' reproducibility, and (3) assess the reproducibility of CT lobar volumes. Materials and Methods: CT imaging and body plethysmography were performed at baseline and after a 9-month interval in multicenter emphysema treatment trials. Lung volumes were measured at total lung capacity (TLC) and at residual volume (RV). Lobar volumes were measured on CT imaging using a semiautomated technique. The correlations between CT and PFT volumes were computed for 486 subjects at baseline. Reproducibility was assessed in terms of the intraclass correlation coefficient (ICC) for 126 subjects from the control group at TLC and 120 subjects at RV. Results: Correlations between CT and PFT lung volumes were 0.86 at TLC and 0.67 at RV. At TLC, the ICCs were 0.943 for CT imaging and 0.814 for PFT. At RV, the ICCs were 0.886 for CT imaging and 0.683 for PFT. CT lobar volumes showed good reproducibility (all P values < .05). Conclusion: CT lung and lobar volume measurements could be captured in a multicenter trial setting with high reproducibility and were highly correlated with those obtained on PFT. CT imaging showed significantly better reproducibility than PFT between interval lung volume measurements, offering the potential for designing emphysema treatment trials involving fewer subjects.
引用
收藏
页码:316 / 322
页数:7
相关论文
共 28 条
[1]
[Anonymous], Clinical trial to evaluate the efficacy and safety of Fesoterodine in comparison to Tolterodine for overactive bladder (OAB) [12 -week, randomized, double-blind, doubledummy,placebo-controlled, parallel -group, multicenter trial to evaluate the efficacy and safety of fesoterodine in comparison to tolterodine ER in patients with overactive bladder (OAB)
[2]
Automated lung segmentation for thoracic CT: Impact on computer-aided diagnosis [J].
Armato, SG ;
Sensakovic, WF .
ACADEMIC RADIOLOGY, 2004, 11 (09) :1011-1021
[3]
Patients with emphysema: Quantitative CT analysis before and after lung volume reduction surgery - Work in progress [J].
Bae, KT ;
Slone, RM ;
Gierada, DS ;
Yusen, RD ;
Cooper, JD .
RADIOLOGY, 1997, 203 (03) :705-714
[4]
MEASUREMENT OF THE VOLUME OF GAS IN THE GASTROINTESTINAL TRACT - VALUES IN NORMAL SUBJECTS AND AMBULATORY PATIENTS [J].
BEDELL, GN ;
MARSHALL, R ;
DUBOIS, AB ;
HARRIS, JH .
JOURNAL OF CLINICAL INVESTIGATION, 1956, 35 (03) :336-345
[5]
Method for segmenting chest CT image data using an anatomical model: Preliminary results [J].
Brown, MS ;
McNitt-Gray, MF ;
Mankovich, NJ ;
Goldin, JG ;
Hiller, J ;
Wilson, LS ;
Aberle, DR .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 1997, 16 (06) :828-839
[6]
Automated measurement of single and total lung volume from CT [J].
Brown, MS ;
McNitt-Gray, MF ;
Goldin, JG ;
Greaser, LE ;
Hayward, UM ;
Sayre, JW ;
Arid, MK ;
Aberle, DR .
JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 1999, 23 (04) :632-640
[7]
Knowledge-based segmentation of thoracic computed tomography images for assessment of split lung function [J].
Brown, MS ;
Goldin, JG ;
McNitt-Gray, MF ;
Greaser, LE ;
Sapra, A ;
Li, KT ;
Sayre, JW ;
Martin, K ;
Aberle, DR .
MEDICAL PHYSICS, 2000, 27 (03) :592-598
[8]
Computed tomography assessment of lung volume changes after bronchial valve treatment [J].
Coxson, H. O. ;
Fauerbach, P. V. Nasute ;
Storness-Bliss, C. ;
Mueller, N. L. ;
Cogswell, S. ;
Dillard, D. H. ;
Finger, C. L. ;
Springmeyer, S. C. .
EUROPEAN RESPIRATORY JOURNAL, 2008, 32 (06) :1443-1450
[9]
MEASUREMENT OF LUNG EXPANSION WITH COMPUTED-TOMOGRAPHY AND COMPARISON WITH QUANTITATIVE HISTOLOGY [J].
COXSON, HO ;
MAYO, JR ;
BEHZAD, H ;
MOORE, BJ ;
VERBURGT, LM ;
STAPLES, CA ;
PARE, PD ;
HOGG, JC .
JOURNAL OF APPLIED PHYSIOLOGY, 1995, 79 (05) :1525-1530
[10]
Repeatability of quantitative CT indexes of emphysema in patients evaluated for lung volume reduction surgery [J].
Gierada, DS ;
Yusen, RD ;
Pilgram, TK ;
Crouch, L ;
Slone, RM ;
Bae, KT ;
Lefrak, SS ;
Cooper, JD .
RADIOLOGY, 2001, 220 (02) :448-454