Rupture of a polypropylene suture after aortic operation: A scanning electronic microscopical assessment of potential mechanisms

被引:7
作者
Huber, C
Eckstein, FS
Halbeisen, M
Carrel, TP [1 ]
机构
[1] Univ Hosp Bern, Cardiovasc Surg Clin, CH-3010 Bern, Switzerland
[2] Swiss Fed Labs Mat Testing & Res, St Gallen, Switzerland
关键词
D O I
10.1016/S0003-4975(02)04679-9
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
in surgical procedures, and specifically in cardiovascular operations, the polypropylene monofilament is a widely used suture material. We report an unexplained case of early rupture of a 4-0 polypropylene suture used for the distal anastomosis after replacement of the ascending aorta with a Dacron (Vaskutek, Sulzer Medica, Austin, TX) prosthesis. By trying to identify if intraoperative mechanical damage could have led to the rupture, we obtained impressive results of how different manipulations of suture material using surgical instruments can damage those delicate filaments. Pictures were obtained by light microscopy and electron microscopy. (C) 2003 by The Society of Thoracic Surgeons.
引用
收藏
页码:1318 / 1321
页数:4
相关论文
共 5 条
[1]   Chronic loading of polypropylene sutures: Implications for breakage after carotid endarterectomy [J].
Dobrin, PB .
JOURNAL OF SURGICAL RESEARCH, 1996, 61 (01) :4-10
[2]   SOME MECHANICAL-PROPERTIES OF POLYPROPYLENE SUTURES - RELATIONSHIP TO THE USE OF POLYPROPYLENE IN VASCULAR-SURGERY [J].
DOBRIN, PB .
JOURNAL OF SURGICAL RESEARCH, 1988, 45 (06) :568-573
[3]   END-TO-SIDE VASCULAR ANASTOMOSIS - MODIFIED TECHNIQUE [J].
ELLERTSON, DG ;
MCGOUGH, EC ;
HUGHES, RK .
ANNALS OF THORACIC SURGERY, 1974, 17 (05) :510-512
[4]   MECHANICAL COMPARISON OF 10 SUTURE MATERIALS BEFORE AND AFTER IN-VIVO INCUBATION [J].
GREENWALD, D ;
SHUMWAY, S ;
ALBEAR, P ;
GOTTLIEB, L .
JOURNAL OF SURGICAL RESEARCH, 1994, 56 (04) :372-377
[5]  
Roy J, 1980, Scan Electron Microsc, P203