Stripping potentiometry signal enhancement using mercury film-coated reticulated vitreous carbon working electrode
被引:10
作者:
Armalis, S
论文数: 0引用数: 0
h-index: 0
机构:
Vilnius State Univ, Dept Analyt & Environm Chem, LT-2006 Vilnius, LithuaniaVilnius State Univ, Dept Analyt & Environm Chem, LT-2006 Vilnius, Lithuania
Armalis, S
[1
]
Kubiliene, E
论文数: 0引用数: 0
h-index: 0
机构:
Vilnius State Univ, Dept Analyt & Environm Chem, LT-2006 Vilnius, LithuaniaVilnius State Univ, Dept Analyt & Environm Chem, LT-2006 Vilnius, Lithuania
Kubiliene, E
[1
]
机构:
[1] Vilnius State Univ, Dept Analyt & Environm Chem, LT-2006 Vilnius, Lithuania
The effect of enhancement of stripping potentiometry signals of trace metals by using a mercury film coated reticulated vitreous carbon flow-through electrode is described. The advantage of such a system is the possibility of the efficient in situ removal of oxygen by exhaustive electrolysis in a small electrode volume during the short stop of the flow after the main deposition time. This results in a considerable enhancement of the analytical signals and also a time saving due to elimination of the unnecessary deaeration procedure. (C) 2000 Elsevier Science B.V. All rights reserved.