Shock-wave measurement using a calibrated interferometric fiber-tip sensor

被引:29
作者
Koch, C
Molkenstruck, W
Reibold, R
机构
[1] Laboratory 1.43, Physikalisch-Tech. Bundesanstalt
[2] Physikalisch-Tech. Bundesanstalt, Lab. 1.43, PF 3345
关键词
shock waves; shock-wave measurement; interferometry; fiberoptic sensors; lithotripsy;
D O I
10.1016/S0301-5629(97)00127-0
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
The results of shock-wave measurements using a calibrated fiber-tip sensor based on a Michelson interferometer are presented, A transfer function, obtained by an independent experiment that describes the properties of the sensor system, was used to correct the measured shock-wave data in the Fourier frequency domain, The phase of the transfer function was determined from its amplitude by a fitting procedure using minimum-phase terms, As an example of application: the acoustic output field of an electromagnetic lithotriptor was investigated, and the shock-wave source was reliably characterized, The measured data provide a basis for estimating the hazard to which a patient is exposed during shock-wave treatment and for optimizing a lithotriptor system to produce a sharply localized and effective acoustic field. (C) 1997 World Federation for Ultrasound in Medicine & Biology.
引用
收藏
页码:1259 / 1266
页数:8
相关论文
共 27 条
[1]  
AHRENDT WR, 1951, AUTOMATIC FEEDBACK C, P137
[2]  
Chaussy C.G., 1988, SHOCK WAVE LITHOTRIP, P1
[3]   A SURVEY OF THE ACOUSTIC OUTPUT OF COMMERCIAL EXTRACORPOREAL SHOCK-WAVE LITHOTRIPTERS [J].
COLEMAN, AJ ;
SAUNDERS, JE .
ULTRASOUND IN MEDICINE AND BIOLOGY, 1989, 15 (03) :213-227
[4]  
DAHMEN GP, 1995, STOSSWELLE FORSCHUNG, P175
[5]   THE CURRENT STATUS OF LITHOTRIPSY [J].
DAWSON, C ;
WHITFIELD, HN .
JOURNAL OF MEDICAL ENGINEERING & TECHNOLOGY, 1994, 18 (06) :224-229
[6]   ACOUSTIC ENERGY DETERMINES HEMOGLOBIN RELEASE FROM ERYTHROCYTES BY EXTRACORPOREAL SHOCK-WAVES IN-VITRO [J].
DELIUS, M ;
UEBERLE, F ;
GAMBIHLER, S .
ULTRASOUND IN MEDICINE AND BIOLOGY, 1995, 21 (05) :707-710
[7]   Cell-type-specific response to shock waves of suspended or pelleted cells as analysed by flow cytometry or electrical cell volume determination [J].
Endl, E ;
Steinbach, P ;
Scharfe, J ;
Fickweiler, S ;
Worle, K ;
Hofstadter, F .
ULTRASOUND IN MEDICINE AND BIOLOGY, 1996, 22 (04) :515-525
[8]   COMPLETE LOCAL TUMOR REMISSION AFTER THERAPY WITH EXTRA-CORPOREALLY APPLIED HIGH-ENERGY SHOCK-WAVES (HESW) [J].
GAMARRA, F ;
SPELSBERG, F ;
DELLIAN, M ;
GOETZ, AE .
INTERNATIONAL JOURNAL OF CANCER, 1993, 55 (01) :153-156
[9]  
GRANZ B, 1994, ULTRASON, P1847, DOI 10.1109/ULTSYM.1994.401950
[10]  
GRANZ B, 1992, J STONE DIS, V4, P123