Methods for studying intestinal sensitivity and compliance: In vitro studies of balloons and a barostat

被引:29
作者
Toma, TP [1 ]
Zighelboim, J [1 ]
Phillips, SF [1 ]
Talley, NJ [1 ]
机构
[1] MAYO CLIN & MAYO FDN,GASTROENTEROL RES UNIT,ROCHESTER,MN 55905
关键词
barostat; compliance; gut latex balloons; polyethylene bag;
D O I
10.1111/j.1365-2982.1996.tb00238.x
中图分类号
R57 [消化系及腹部疾病];
学科分类号
摘要
The aim of this study was to compare in vitro various methods for recording intestinal sensitivity and compliance. Relationships between volume and pressure were determined in segments of penrose tubing and pig gut ('artificial intestine') using pressure increments of 2 mmHg (0-24 mmHg). We tested two direct methods of distension of the entire segments (by syringe inflation and the Mayo barostat); we also used three different balloon devices for indirect distension (a 10 cm polyethylene barostat bag, a 10 cm latex condom balloon and a 6 cm latex condom balloon). Maximal distending diameters of the recording systems were measured by injecting from 0 to 160 mL of air. The elastic properties of the balloons were also tested by distensions in air and in rigid tubes. All recording systems accurately detected a lesser compliance of the penrose drain as compared to pig gut. In absolute terms, only the compliance measured with a polyethylene barostat bag distended with a syringe was not different from the compliance of the segment as measured directly. The bellows of our barostat and the latex balloons had significant intrinsic compliances which interfered with the recorded pressure-volume curves. On the other hand, highly compliant plastic bags recorded most faithfully the compliance of artificial gut and that of noncompliant rigid tubes. For comparable volumes of distension, external diameters were larger with the 6 cm latex balloon than with the 10 cm latex balloon or the 10 cm polyethylene barostat balloon. A polyethylene bag distended with a non-compliant air injector (syringe) reflected most accurately the pressure-volume relationships of tubular structures. The different maximal diameters assumed by the three distending devices may explain, in part, why lower Volumes of distension are required to elicit symptoms with smaller distending balloons in vivo.
引用
收藏
页码:19 / 28
页数:10
相关论文
共 23 条
[1]   RECTAL RESERVOIR AND SENSORY FUNCTION STUDIED BY GRADED ISOBARIC DISTENSION IN NORMAL MAN [J].
AKERVALL, S ;
FASTH, S ;
NORDGREN, S ;
ORESLAND, T ;
HULTEN, L .
GUT, 1989, 30 (04) :496-502
[2]   PERCEPTION AND REFLEX RELAXATION OF THE STOMACH IN RESPONSE TO GUT DISTENSION [J].
AZPIROZ, F ;
MALAGELADA, JR .
GASTROENTEROLOGY, 1990, 98 (05) :1193-1198
[3]   GASTRIC TONE MEASURED BY AN ELECTRONIC BAROSTAT IN HEALTH AND POSTSURGICAL GASTROPARESIS [J].
AZPIROZ, F ;
MALAGELADA, JR .
GASTROENTEROLOGY, 1987, 92 (04) :934-943
[4]   PHYSIOLOGICAL VARIATIONS IN CANINE GASTRIC TONE MEASURED BY AN ELECTRONIC BAROSTAT [J].
AZPIROZ, F ;
MALAGELADA, JR .
AMERICAN JOURNAL OF PHYSIOLOGY, 1985, 248 (02) :G229-G237
[5]   VARIATIONS IN MUSCLE TONE OF THE HUMAN RECTUM - RECORDINGS WITH AN ELECTROMECHANICAL BAROSTAT [J].
BELL, AM ;
PEMBERTON, JH ;
HANSON, RB ;
ZINSMEISTER, AR .
AMERICAN JOURNAL OF PHYSIOLOGY, 1991, 260 (01) :G17-G25
[6]  
HACHET T, 1994, GASTROENTEROL CLIN B, V17, P347
[7]  
HAMMER J, 1993, ALIMENT PHARM THERAP, V7, P543
[8]   A SOMATOSTATIN ANALOG INHIBITS AFFERENT PATHWAYS MEDIATING PERCEPTION OF RECTAL DISTENSION [J].
HASLER, WL ;
SOUDAH, HC ;
OWYANG, C .
GASTROENTEROLOGY, 1993, 104 (05) :1390-1397
[9]   INTERINDIVIDUAL AND INTRAINDIVIDUAL VARIATION IN PRESSURE-VOLUME RELATIONS OF THE RECTUM IN NORMAL SUBJECTS AND PATIENTS WITH THE IRRITABLE-BOWEL-SYNDROME [J].
KENDALL, GPN ;
THOMPSON, DG ;
DAY, SJ ;
LENNARDJONES, JE .
GUT, 1990, 31 (09) :1062-1068
[10]   MEASUREMENT OF HUMAN ESOPHAGEAL TONE IN-VIVO [J].
MAYRAND, S ;
DIAMANT, NE .
GASTROENTEROLOGY, 1993, 105 (05) :1411-1420