Evaluation of monolithic and sub 2 mm particle packed columns for the rapid screening for illicit drugs - application to the determination of drug contamination on Irish euro banknotes

被引:31
作者
Bones, Jonathan [1 ]
Macka, Mirek [1 ]
Paull, Brett [1 ]
机构
[1] Dublin City Univ, Sch Chem Sci, Natl Ctr Sensor Res, Dublin 9, Ireland
关键词
D O I
10.1039/b615669j
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A study comparing recently available 100 x 3 mm id, 200 x 3 mm id monolithic reversed-phase columns with a 50 x 2.1 mm id, 1.8 mu m particle packed reversed-phase columns was carried out to determine the most efficient approach (using traditional van Deemter analysis and a modern kinetic plot approach) for the rapid screening of samples for 16 illicit drugs and associated metabolites. A plot of column backpressure versus plate number (N) showed a significant advantage of using the monolithic phases, with the 20 cm monolithic column exhibiting a maximum 15 000 plates at a column backpressure of approximate to 70 bar, compared to approximate to 7000 plates at 150 bar for the 5 cm 1.8 mm particle packed column. Optimum linear velocities were found to be 0.40 mm s(-1), 0.52 mm s(-1) and 0.98 mm s(-1) for the three above columns, respectively. The 20 cm monolithic column was subsequently applied to the separation and determination of illicit drug contamination on Irish euro banknotes, using methanol extraction followed by LC-MS/MS. Method performance data showed that the new LC-MS/MS method was significantly more sensitive than previous GC-MS/MS based methods for this application, with detection limits in the pg note(-1) region, based upon a 20 mu L standard injection. All of the notes examined tested positive for trace quantities of cocaine, with benzoylecgonine detected on 12 of the 45 notes sampled. Traces of heroin were also detected on three of the 45 notes.
引用
收藏
页码:208 / 217
页数:10
相关论文
共 34 条
[1]  
BURKEKENNEDY E, 2004, IRISH TIMES 0729
[2]   Applications of silica-based monolithic HPLC columns [J].
Cabrera, K .
JOURNAL OF SEPARATION SCIENCE, 2004, 27 (10-11) :843-852
[3]   The distribution of controlled drugs on banknotes via counting machines [J].
Carter, JF ;
Sleeman, R ;
Parry, J .
FORENSIC SCIENCE INTERNATIONAL, 2003, 132 (02) :106-112
[4]   HIGH-SPEED HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY OF PEPTIDES AND PROTEINS [J].
CHEN, H ;
HORVATH, C .
JOURNAL OF CHROMATOGRAPHY A, 1995, 705 (01) :3-20
[5]   Influence of frictional heating on temperature gradients in ultra-high-pressure liquid chromatography on 2.1 mm ID columns [J].
de Villiers, A ;
Lauer, H ;
Szucs, R ;
Goodall, S ;
Sandra, P .
JOURNAL OF CHROMATOGRAPHY A, 2006, 1113 (1-2) :84-91
[6]  
Desmet G, 2005, LC GC EUR, V18, P403
[7]   Geometry-independent plate height representation methods for the direct comparison of the kinetic performance of LC supports with a different size or morphology [J].
Desmet, G ;
Clicq, D ;
Gzil, P .
ANALYTICAL CHEMISTRY, 2005, 77 (13) :4058-4070
[8]  
DESMET G, KINETIC PLOT METHOD
[9]   Rapid comparison of diacetylmorphine on banknotes by tandem mass spectrometry [J].
Ebejer, KA ;
Brereton, RG ;
Carter, JF ;
Ollerton, SL ;
Sleeman, R .
RAPID COMMUNICATIONS IN MASS SPECTROMETRY, 2005, 19 (15) :2137-2143
[10]  
EBEJER KA, 2006, IN PRESS FORENSIC SC, DOI DOI 10.1016/J.FORENSIC.2006.06.039