Dynamic response of a hot-wire anemometer. Part II: A flush-mounted hot-wire and hot-film probes for wall shear stress measurements

被引:42
作者
Chew, YT [1 ]
Khoo, BC [1 ]
Lim, CP [1 ]
Teo, CJ [1 ]
机构
[1] Natl Univ Singapore, Dept Mech & Prod Engn, Singapore 119260, Singapore
关键词
dynamic response; flush-mounted hot wire; hot-film wall shear stress gauge; wall shear stress fluctuations;
D O I
10.1088/0957-0233/9/5/006
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
The present study continues the work described in Part I of this paper in utilizing a specifically designed apparatus for generating a known near-wall fluctuating flow field for the purpose of quantifying the dynamic response of a flush-mounted hot-element wall shear stress gauge. Two different types of wall shear stress gauges were tested: a flush-mounted hot wire in contact with the wall substrate and commercially available quartz substrate hot-film gauges with different thicknesses of quartz coating, It is found that the flush-mounted hot-wire gauge has a much higher dynamic response than the quartz substrate gauges, although it is still lower than the marginally elevated hot-wire configuration, This may suggest the possible use of a marginally elevated hot wire as a wall shear stress gauge to ensure sufficient responsiveness for time-resolved wall shear stress measurements.
引用
收藏
页码:764 / 778
页数:15
相关论文
共 23 条
[1]   MODIFIED GAUGE FOR TIME-RESOLVED SKIN-FRICTION MEASUREMENTS [J].
AJAGU, CO ;
LIBBY, PA ;
LARUE, JC .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1982, 53 (12) :1920-1926
[2]   THE FLUCTUATING WALL-SHEAR STRESS AND THE VELOCITY-FIELD IN THE VISCOUS SUBLAYER [J].
ALFREDSSON, PH ;
JOHANSSON, AV ;
HARITONIDIS, JH ;
ECKELMANN, H .
PHYSICS OF FLUIDS, 1988, 31 (05) :1026-1033
[3]   MEASUREMENT OF SHEAR-STRESS NEAR A STAGNATION POINT [J].
BAINES, WD ;
KEFFER, JF .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1976, 47 (04) :440-442
[4]   An investigation of wall effects on hot-wire measurements using a bent sublayer probe [J].
Chew, YT ;
Khoo, BC ;
Li, GL .
MEASUREMENT SCIENCE AND TECHNOLOGY, 1998, 9 (01) :67-85
[5]   A TIME-RESOLVED HOT-WIRE SHEAR-STRESS PROBE FOR TURBULENT-FLOW - USE OF LAMINAR-FLOW CALIBRATION [J].
CHEW, YT ;
KHOO, BC ;
LI, GL .
EXPERIMENTS IN FLUIDS, 1994, 17 (1-2) :75-83
[6]   RESPONSE OF HOT-ELEMENT WALL SHEAR-STRESS GAUGES IN UNSTEADY TURBULENT FLOWS [J].
COOK, WJ .
AIAA JOURNAL, 1994, 32 (07) :1464-1471
[7]   RESPONSE OF HOT-ELEMENT WALL SHEAR-STRESS GAUGES IN LAMINAR OSCILLATING FLOWS [J].
COOK, WJ ;
GIDDINGS, TA ;
MURPHY, JD .
AIAA JOURNAL, 1988, 26 (06) :706-713
[8]   FREQUENCY-RESPONSE AND ELECTRONIC TESTING FOR CONSTANT-TEMPERATURE HOT-WIRE ANEMOMETERS [J].
FREYMUTH, P .
JOURNAL OF PHYSICS E-SCIENTIFIC INSTRUMENTS, 1977, 10 (07) :705-710
[9]   CALCULATION OF SQUARE-WAVE TEST FOR FREQUENCY OPTIMIZED HOT-FILM ANEMOMETERS [J].
FREYMUTH, P .
JOURNAL OF PHYSICS E-SCIENTIFIC INSTRUMENTS, 1981, 14 (02) :238-240
[10]  
Freymuth P., 1977, TSI Q, VIII, P5