Spiral interpolation algorithms for multislice spiral CT - Part II: Measurement and evaluation of slice sensitivity profiles and noise at a clinical multislice system

被引:32
作者
Fuchs, T [1 ]
Krause, J
Schaller, S
Flohr, T
Kalender, WA
机构
[1] Univ Erlangen Nurnberg, Inst Med Phys, D-91054 Erlangen, Germany
[2] Siemens AG, Med Engn Grp, D-91301 Forchheim, Germany
关键词
computed tomography (CT); image quality (IQ); multirow detector system; multislice spiral CT (MSCT); noise; slice; sensitivity profile (SSP);
D O I
10.1109/42.887833
中图分类号
TP39 [计算机的应用];
学科分类号
081203 ; 0835 ;
摘要
The recently introduced multislice data acquisition for computed tomography (CT) is based on multirow detector design, increased rotation speed, and advanced z-interpolation and z-filtering algorithms. We evaluated slice sensitivity profiles (SSPs) and noise of a clinical multislice spiral CT (MSCT) scanner with M = 4 simultaneously acquired slices and adaptive axial interpolator (AAI) reconstruction software [11]. SSPs were measured with a small gold disk of 50 mum thickness and 2-mm diameter located at the center of rotation (COR) and 100 mm off center. The standard deviation of CT values within a 20-cm water phantom was used as a measure of image noise. With a detector slice collimation of S = 1.0 mm, we varied spiral pitch p from 0.25 to 2.0 in steps of 0.025. Nominal reconstructed slice thicknesses were 1.25, 1.5, and 2.0 mm. For all possible pitch values, we found the full-width at half maximum (FWHM) of the respective sensitivity profile at the COR equivalent to the selected nominal slice thickness. The profiles at 100 mm off center are broadened less than 7% on the average compared with the FWHM at the COR, In addition, variation of the full-width at tenth maximum (FWTM) at the COR was below 3.0% for p less than or equal to 1.75. Within this range, image noise varied less than 10% with respect to the mean noise level. The slight increase in measured slice-width above p = 1.75 for nominal slice-widths of 1.25 and 1.50 mm is accompanied by a decrease of noise according to the inverse square root relationship. The MSCT system that we scrutinized provides reconstructed slice-widths and image noise, which can be regarded as constant within a wide range of table speeds. With respect to this, MSCT is superior to single-slice spiral CT. These facts can be made use of when defining and optimizing clinical protocols: the spiral pitch can be selected almost freely, and scan protocols can follow the diagnostic requirements without technical restrictions. In summary, MSCT offers constant image quality while scan times are reduced drastically. Volume scans with three-dimensional (3-D) isotropic resolution are routinely feasible for complete anatomical regions.
引用
收藏
页码:835 / 847
页数:13
相关论文
共 18 条
[1]  
[Anonymous], 2000, COMPUTED TOMOGRAPHY
[2]   Helical CT reconstruction with longitudinal filtration [J].
Hu, H ;
Shen, Y .
MEDICAL PHYSICS, 1998, 25 (11) :2130-2138
[3]   Multi-slice helical CT: Scan and reconstruction [J].
Hu, H .
MEDICAL PHYSICS, 1999, 26 (01) :5-18
[4]  
International Electrotechnical Commission (IEC), 1999, 60601244 IEC
[5]  
Kachelriess M, 1999, RADIOLOGY, V213P, P283
[6]   ECG-correlated image reconstruction from subsecond multi-slice spiral CT scans of the heart [J].
Kachelriess, M ;
Ulzheimer, S ;
Kalender, WA .
MEDICAL PHYSICS, 2000, 27 (08) :1881-1902
[7]   SPIRAL VOLUMETRIC CT WITH SINGLE-BREATH-HOLD TECHNIQUE, CONTINUOUS TRANSPORT, AND CONTINUOUS SCANNER ROTATION [J].
KALENDER, WA ;
SEISSLER, W ;
KLOTZ, E ;
VOCK, P .
RADIOLOGY, 1990, 176 (01) :181-183
[8]   A COMPARISON OF CONVENTIONAL AND SPIRAL CT - AN EXPERIMENTAL-STUDY ON THE DETECTION OF SPHERICAL LESIONS [J].
KALENDER, WA ;
POLACIN, A ;
SUSS, C .
JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 1994, 18 (02) :167-176
[9]  
KLINGENBECKREGN K, IN PRESS EUR J RADIO
[10]   Performance evaluation of a multi-slice CT system [J].
McCollough, CH ;
Zink, FE .
MEDICAL PHYSICS, 1999, 26 (11) :2223-2230