摘要
A reversed-phase (RP) high-performance liquid chromatographic (HPLC) method with fluorescence detection allowing the sensitive and specific quantification of BAY 12-8039, a new antimicrobially active 8-methoxyquinolone, in biological fluids is described, The method is compared to a microbiological assay (bioassay) based on B. subtilis test strain with a limit of quantification of approximately 60 mu g/l. Following dilution and centrifugation, plasma, saliva or urine supernatant is directly injected onto the HPLC system. Concentrations down to a limit of quantification of 2.5 mu g/l can be quantified in plasma, saliva and urine. Data on recovery, accuracy and precision of the method throughout the whole working range as well as results on stability of the analyte are presented. The concentration data are correlated with results from the bioassay. BAY 12-8039 is stable in plasma after repeated freeze-thaw cycles and following storage at -20 degrees C for at least 12 months. The results of HPLC measurements excellently agree with bioassay data indicating the relevance of the method as a tool in clinical development to answer pharmacokinetic questions related to antimicrobial activity. The method was applied to human plasma, saliva and urine from subjects after a single oral dose of 400 mg of BAY 12-8039. (C) 1997 Elsevier Science B.V.
引用
收藏
页码:163 / 174
页数:12
相关论文
共 28 条
[1]   A strategy for validation of bioanalytical methods [J].
Braggio, S ;
Barnaby, RJ ;
Grossi, P ;
Cugola, M .
JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 1996, 14 (04) :375-388
[2]   A COMPARATIVE-STUDY OF LARGE VOLUME INJECTION TECHNIQUES FOR MICROBORE COLUMNS IN HPLC [J].
CLAESSENS, HA ;
KUYKEN, MAJ .
CHROMATOGRAPHIA, 1987, 23 (05) :331-336
[3]   In vitro activity of BAY 12-8039, a new 8-methoxyquinolone [J].
Dalhoff, A ;
Petersen, U ;
Endermann, R .
CHEMOTHERAPY, 1996, 42 (06) :410-425
[4]  
*FDA, 1984, DRAFT GUID STAB STUD
[5]   GENERAL-METHOD VALIDATION GUIDELINES FOR PHARMACEUTICAL SAMPLES [J].
INMAN, EL ;
FRISCHMANN, JK ;
JIMENEZ, PJ ;
WINKEL, GD ;
PERSINGER, ML ;
RUTHERFORD, BS .
JOURNAL OF CHROMATOGRAPHIC SCIENCE, 1987, 25 (06) :252-256
[6]   PRECISION, ACCURACY, AND DATA ACCEPTANCE CRITERIA IN BIOPHARMACEUTICAL ANALYSIS [J].
KARNES, HT ;
MARCH, C .
PHARMACEUTICAL RESEARCH, 1993, 10 (10) :1420-1426
[7]   VALIDATION OF BIOANALYTICAL METHODS [J].
KARNES, HT ;
SHIU, G ;
SHAH, VP .
PHARMACEUTICAL RESEARCH, 1991, 8 (04) :421-426
[8]  
KOHLSDORFER C, 1994, BAY 128036 ANAL METH
[9]  
KUBITZA D, 1996, P 36 ICAAC NEW ORL, P104
[10]   COMPARATIVE-STUDY OF ON-COLUMN FOCUSING APPLIED TO MICRO-LC AND CONVENTIONAL LC [J].
LING, BL ;
BAEYENS, W ;
DEWAELE, C .
JOURNAL OF MICROCOLUMN SEPARATIONS, 1992, 4 (01) :17-22