In vitro calibration and validation of a digital subtraction radiography system using scanned images

被引:13
作者
Woo, BMS
Zee, KY
Chan, FHY
Corbet, EF
机构
[1] Univ Hong Kong, Dept Elect & Elect Engn, Hong Kong, Hong Kong, Peoples R China
[2] Univ Hong Kong, Fac Dent, Prince Philip Dent Hosp 3E, Hong Kong, Hong Kong, Peoples R China
关键词
computer-assisted densitometric image analysis; digital subtraction radiography; radiographs; scanned images;
D O I
10.1034/j.1600-051X.2003.00236.x
中图分类号
R78 [口腔科学];
学科分类号
1003 [口腔医学];
摘要
Objectives: To calibrate and validate a digital subtraction radiography system using scanned images for quantification of alveolar bone changes by means of computer-assisted densitometric image analysis (CADIA) in vitro. Materials and Methods: Noise levels were determined using 10 standardized periapical radiographs of the same lower molar region in a human dry skull. For validation of the system, radiographs were taken before and after bovine bone particles in measures with increments of 2 mg weighing from 2 to 20 mg were added into each socket of three dry skulls. Radiographs were developed and scanned into a computer with a flatbed scanner. After digitization, the images were subjected to alignment, normalization and subtraction. Appropriate regions of interest (ROIs) were selected and their CADIA values were calculated for the determination of noise levels, and correlations between the CADIA values and the actual bone mass were performed. Results: When the threshold value was 7, the percentage of pixels deviating from the set threshold value was small (0-11.3%). There were statistically significant correlations between the actual bone mass and the CADIA value for anterior sockets (p<0.001, r(2)=0.89) and posterior sockets (p<0.001, r (2) =0.9). For pooled data of both anterior and posterior sockets, the correlation was also statistically significant (p< 0.001, r(2)=0.88). Conclusions: A high and statistically significant correlation between the actual bone mass and CADIA value was obtained, which suggests that the system could be suitable for the detection of small alveolar bone changes.
引用
收藏
页码:114 / 118
页数:5
相关论文
共 22 条
[1]
COMPUTER-ASSISTED DENSITOMETRIC IMAGE-ANALYSIS FOR THE QUANTITATION OF RADIOGRAPHIC ALVEOLAR BONE CHANGES [J].
BRAEGGER, U ;
LITCH, J ;
PASQUALI, L ;
KORNMAN, KS .
JOURNAL OF PERIODONTAL RESEARCH, 1987, 22 (03) :227-229
[2]
COMPUTER-ASSISTED DENSITOMETRIC IMAGE-ANALYSIS IN PERIODONTAL RADIOGRAPHY - A METHODOLOGICAL STUDY [J].
BRAGGER, D ;
PASQUALI, L ;
RYLANDER, H ;
CARNES, D ;
KORNMAN, KS .
JOURNAL OF CLINICAL PERIODONTOLOGY, 1988, 15 (01) :27-37
[3]
Brägger U, 1998, J PERIODONTOL, V69, P967
[4]
DIGITAL IMAGING IN PERIODONTAL RADIOGRAPHY - A REVIEW [J].
BRAGGER, U .
JOURNAL OF CLINICAL PERIODONTOLOGY, 1988, 15 (09) :551-557
[5]
Quantitative digital subtraction radiography for the determination of small changes in bone thickness - An in vitro study [J].
Christgau, M ;
Hiller, KA ;
Schmalz, G ;
Kolbeck, C ;
Wenzel, A .
ORAL SURGERY ORAL MEDICINE ORAL PATHOLOGY ORAL RADIOLOGY AND ENDODONTOLOGY, 1998, 85 (04) :462-472
[6]
Dreyer W P, 1993, Int Dent J, V43, P557
[7]
DIGITAL IMAGE-PROCESSING .2. IN-VITRO QUANTITATIVE-EVALUATION OF SOFT AND HARD PERI-IMPLANT TISSUE CHANGES [J].
FOURMOUSIS, I ;
BRAGGER, U ;
BURGIN, W ;
TONETTI, M ;
LANG, NP .
CLINICAL ORAL IMPLANTS RESEARCH, 1994, 5 (02) :105-114
[8]
DIGITAL IMAGE-PROCESSING .1. EVALUATION OF GRAY-LEVEL CORRECTION METHODS IN-VITRO [J].
FOURMOUSIS, I ;
BRAGGER, U ;
BURGIN, W ;
TONETTI, M ;
LANG, NP .
CLINICAL ORAL IMPLANTS RESEARCH, 1994, 5 (01) :37-47
[9]
THE RELATIONSHIP BETWEEN ATTACHMENT LEVEL LOSS AND ALVEOLAR BONE LOSS [J].
GOODSON, JM ;
HAFFAJEE, AD ;
SOCRANSKY, SS .
JOURNAL OF CLINICAL PERIODONTOLOGY, 1984, 11 (05) :348-359
[10]
A DIGITAL SUBTRACTION TECHNIQUE FOR DENTAL RADIOGRAPHY [J].
GRONDAHL, HG ;
GRONDAHL, K ;
WEBBER, RL .
ORAL SURGERY ORAL MEDICINE ORAL PATHOLOGY ORAL RADIOLOGY AND ENDODONTOLOGY, 1983, 55 (01) :96-102