Two-point fluorescence detection and automated fraction collection applied to constant denaturant capillary electrophoresis

被引:22
作者
Ekstrom, P
Wasserkort, R
Minarik, M
Foret, F
Thilly, WG
机构
[1] MIT, Cambridge, MA 02139 USA
[2] Northeastern Univ, Boston, MA 02115 USA
关键词
D O I
10.2144/00293rr01
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Constant denaturant capillary electrophoresis (CDCE) has been shown to be a sensitive method to detect point mutations in DNA sequences of 100-bp lengths. Here, we report a significant modifications for the instrumental setup that allows a highly accurate prediction of the elution time of DNA fragments from the capillary and an efficient collection of separated fractions. Fluorescently labeled DNA fragments of TP53 exon 8 wild-type and two mutants (base pair number 14480 and 14525) are detected at two separate points of the same capillary. This permits the precise calculation of the fragment velocity after separation in the heated one because, at room temperature, all DNA fragments of the same length have the same velocity. Such precision permits the selective collection of separated fragments using an automated fraction collector for additional CDCE analysis or sequencing. Also, the two-point detection allows one to rapidly distinguish between double-stranded and single-stranded DNA fragments of the same length, a process that cannot be achieved with a one-point detection system alone. Both modifications greatly improve the procedure to detect novel mutations by means of CDCE.
引用
收藏
页码:582 / +
页数:6
相关论文
共 23 条
[1]   Fidelity and mutational spectrum of Pfu DNA polymerase on a human mitochondrial DNA sequence [J].
Andre, P ;
Kim, A ;
Khrapko, K ;
Thilly, WG .
GENOME RESEARCH, 1997, 7 (08) :843-852
[2]   Mutation analyses of KRAS exon 1 comparing three different techniques:: temporal temperature gradient electrophoresis, constant denaturant capillary electrophoresis and allele specific polymerase chain reaction [J].
Bjorheim, J ;
Lystad, S ;
Lindblom, A ;
Kressner, U ;
Westring, S ;
Wahlberg, S ;
Lindmark, G ;
Gaudernack, G ;
Ekstrom, P ;
Roe, J ;
Thilly, WG ;
Borresen-Dale, AL .
MUTATION RESEARCH-FUNDAMENTAL AND MOLECULAR MECHANISMS OF MUTAGENESIS, 1998, 403 (1-2) :103-112
[3]  
BORRESEN AL, 1992, CANCER RES, V52, P3234
[4]   CONSTANT DENATURANT GEL-ELECTROPHORESIS AS A RAPID SCREENING TECHNIQUE FOR P53 MUTATIONS [J].
BORRESEN, AL ;
HOVIG, E ;
SMITHSORENSEN, B ;
MALKIN, D ;
LYSTAD, S ;
ANDERSEN, TI ;
NESLAND, JM ;
ISSELBACHER, KJ ;
FRIEND, SH .
PROCEEDINGS OF THE NATIONAL ACADEMY OF SCIENCES OF THE UNITED STATES OF AMERICA, 1991, 88 (19) :8405-8409
[5]   Detection of low-frequency mutations in exon 8 of the TP53 gene by constant denaturant capillary electrophoresis (CDCE) [J].
Ekstrom, PO ;
Borresen-Dayle, AL ;
Qvist, H ;
Giercksky, KE ;
Thilly, WG .
BIOTECHNIQUES, 1999, 27 (01) :128-+
[6]   DNA FRAGMENTS DIFFERING BY SINGLE BASE-PAIR SUBSTITUTIONS ARE SEPARATED IN DENATURING GRADIENT GELS - CORRESPONDENCE WITH MELTING THEORY [J].
FISCHER, SG ;
LERMAN, LS .
PROCEEDINGS OF THE NATIONAL ACADEMY OF SCIENCES OF THE UNITED STATES OF AMERICA-BIOLOGICAL SCIENCES, 1983, 80 (06) :1579-1583
[7]   A SIMPLE METHOD FOR IDENTIFICATION OF POINT MUTATIONS USING DENATURING GRADIENT GEL-ELECTROPHORESIS [J].
GULDBERG, P ;
GUTTLER, F .
NUCLEIC ACIDS RESEARCH, 1993, 21 (09) :2261-2262
[8]   HIGH-PERFORMANCE ELECTROPHORESIS - ELIMINATION OF ELECTROENDOSMOSIS AND SOLUTE ADSORPTION [J].
HJERTEN, S .
JOURNAL OF CHROMATOGRAPHY, 1985, 347 (02) :191-198
[9]   CONSTANT DENATURANT GEL-ELECTROPHORESIS, A MODIFICATION OF DENATURING GRADIENT GEL-ELECTROPHORESIS, IN MUTATION DETECTION [J].
HOVIG, E ;
SMITHSORENSEN, B ;
BROGGER, A ;
BORRESEN, AL .
MUTATION RESEARCH, 1991, 262 (01) :63-71
[10]   Efficiency of separation of DNA mutations by constant denaturant capillary electrophoresis is controlled by the kinetics of DNA melting equilibrium [J].
Khrapko, K ;
Coller, H ;
Thilly, W .
ELECTROPHORESIS, 1996, 17 (12) :1867-1874