Finite Element Analysis of Hepatic Radiofrequency Ablation Probes using Temperature-Dependent Electrical Conductivity

被引:110
作者
Chang, Isaac [1 ]
机构
[1] US FDA, Off Sci & Technol, Ctr Devices & Radiol Hlth, Rockville, MD 20857 USA
关键词
Radiofrequency Ablation; Tissue Perfusion; Ablation Probe; Specific Absorption Rate; Proximal Edge;
D O I
10.1186/1475-925X-2-12
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Background: Few finite element models (FEM) have been developed to describe the electric field, specific absorption rate (SAR), and the temperature distribution surrounding hepatic radiofrequency ablation probes. To date, a coupled finite element model that accounts for the temperature-dependent electrical conductivity changes has not been developed for ablation type devices. While it is widely acknowledged that accounting for temperature dependent phenomena may affect the outcome of these models, the effect has not been assessed. Methods: The results of four finite element models are compared: constant electrical conductivity without tissue perfusion, temperature-dependent conductivity without tissue perfusion, constant electrical conductivity with tissue perfusion, and temperature-dependent conductivity with tissue perfusion. Results: The data demonstrate that significant errors are generated when constant electrical conductivity is assumed in coupled electrical-heat transfer problems that operate at high temperatures. These errors appear to be closely related to the temperature at which the ablation device operates and not to the amount of power applied by the device or the state of tissue perfusion. Conclusion: Accounting for temperature-dependent phenomena may be critically important in the safe operation of radiofrequency ablation device that operate near 100 degrees C.
引用
收藏
页数:18
相关论文
共 26 条
[1]   CDRH RF PHANTOM FOR HYPERTHERMIA SYSTEMS EVALUATIONS [J].
ALLEN, S ;
KANTOR, G ;
BASSEN, H ;
RUGGERA, P .
INTERNATIONAL JOURNAL OF HYPERTHERMIA, 1988, 4 (01) :17-23
[2]   A heat transfer model of thermal balloon endometrial ablation [J].
Baldwin, SA ;
Pelman, A ;
Bert, JL .
ANNALS OF BIOMEDICAL ENGINEERING, 2001, 29 (11) :1009-1018
[3]  
BASSEN HI, 1981, J MICROWAVE POWER EE, V16, P215
[4]  
Braunwald E, 1987, HARRISONS PRINCIPLES, P418
[5]   Precision test apparatus for evaluating the heating pattern of radiofrequency ablation devices [J].
Chang, I ;
Beard, B .
MEDICAL ENGINEERING & PHYSICS, 2002, 24 (09) :633-640
[6]  
Curcie D J, 1995, Technol Health Care, V3, P111
[7]   Interstitial bipolar RF-thermotherapy (RFITT) - Therapy planning by computer simulation and MRI-monitoring - A new concept for minimally invasive procedures [J].
Desinger, K ;
Stein, T ;
Muller, G ;
Mack, M ;
Vogl, TJ .
SURGICAL APPLICATIONS OF ENERGY, PROCEEDINGS OF, 1998, 3249 :147-159
[8]  
Duck F. A., 1990, Physical properties of tissues, P167, DOI [10.1016/B978-0-12-222800-1.50010-3, DOI 10.1016/B978-0-12-222800-1.50010-3]
[9]   MYOCARDIAL ELECTRICAL-IMPEDANCE MAPPING OF ISCHEMIC SHEEP HEARTS AND HEALING ANEURYSMS [J].
FALLERT, MA ;
MIROTZNIK, MS ;
DOWNING, SW ;
SAVAGE, EB ;
FOSTER, KR ;
JOSEPHSON, ME ;
BOGEN, DK .
CIRCULATION, 1993, 87 (01) :199-207
[10]   The dielectric properties of biological tissues .1. Literature survey [J].
Gabriel, C ;
Gabriel, S ;
Corthout, E .
PHYSICS IN MEDICINE AND BIOLOGY, 1996, 41 (11) :2231-2249