seroconversion;
HIV combined antigen and antibody assay;
p24;
antigen;
D O I:
10.1016/j.jviromet.2004.08.018
中图分类号:
Q5 [生物化学];
学科分类号:
071010 ;
081704 ;
摘要:
In this study, we evaluated the performance of six HIV combined p24 antigen and antibody (Ag/Ab) assays versus two third-generation anti-HIV antibody assays. The assays were evaluated using p24 antigen panel of 31 HIV-1 subtypes (n= 124), 25 HIV-1 seroconversion panels (n = 176) HIV-1 antibody positive samples including group M subtypes and group 0 (n = 559), HIV-2 positive samples (n = 110), and unselected HIV negative samples from four French private laboratories (n = 1005). The results showed that overall HIV combined Ag/Ab assays present better performance. when compared to antibody-only assays. However, some differences were observed in the sensitivity of the six HIV combined Ag/Ab assays evaluated. The AxSYM and Murex Combo assays had the best sensitivity score in this study and reduced the window period by 2.0-2.35 days relative to antibody only assays and 1-2.17 days relative to the other combined Ag/Ab assays. Among combined HIV Ag/Ab assays, Genscreen Plus and AxSYM Combo presented the highest specificity, with 99.9% and 99.8%, respectively. (C) 2004 Elsevier B.V. All rights reserved.