Failed Metal-on-Metal Hip Arthroplasties: A Spectrum of Clinical Presentations and Operative Findings

被引:105
作者
Browne, James A. [1 ]
Bechtold, C. Dustin [1 ]
Berry, Daniel J. [1 ]
Hanssen, Arlen D. [1 ]
Lewallen, David G. [1 ]
机构
[1] Mayo Clin, Rochester, MN 55905 USA
关键词
FAILURE; BEARINGS; HYPERSENSITIVITY; RESURFACINGS; PSEUDOTUMOR; REPLACEMENT; NECK;
D O I
10.1007/s11999-010-1419-0
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
100224 [整形外科学];
摘要
A number of recent reports have described novel failure mechanisms of metal-on-metal bearings in total and resurfacing hip arthroplasty. Hip arthroplasties with metal-on-metal articulations are also subject to the traditional methods of failure seen with different bearing couples. There is currently little information in the literature to help guide timely clinical evaluation and management of these patients. We therefore describe the (1) clinical presentations; (2) reasons for failure; (3) operative findings; and (4) histologic findings in patients with failed metal-on-metal hip arthroplasties. We retrospectively identified all 37 patients (37 hips) with metal on metal total hip or resurfacing arthroplasties who underwent revision over the past 3 years at our institution. Relevant clinical, radiographic, laboratory, intraoperative, and histopathologic findings were analyzed for all patients. Of the 37 patients, 10 were revised for presumed hypersensitivity specific to the metal-on-metal articulation. This group included eight patients with tissue histology confirming chronic inflammation with lymphocytic infiltration, eight with aseptic loosening of a monoblock screwless uncemented acetabular component, two with iliopsoas impingement associated with a large-diameter femoral head, and three with femoral neck fracture after resurfacing arthroplasty; the remainder of the patients were revised for infection, instability, component malposition, and periprosthetic fracture. Increased awareness of the modes of failure will bring to light the potential complications particular to metal-on-metal articulations while placing these complications into the context of failures associated with all hip arthroplasties. This novel clinical information should be valuable for the practicing surgeon faced with this patient population. Level IV, therapeutic study. See Guidelines for Authors for a complete description of levels of evidence.
引用
收藏
页码:2313 / 2320
页数:8
相关论文
共 21 条
[1]
Fracture of the neck of the femur after surface arthroplasty of the hip [J].
Amstutz, HC ;
Campbell, PA ;
Le Duff, MJ .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2004, 86A (09) :1874-1877
[2]
AROUKATOS P, 2009, CLIN ORTHOP REL 1218
[3]
BIANT LC, 2009, J ARTHROPLASTY 0223
[4]
The Epidemiology of Bearing Surface Usage in Total Hip Arthroplasty in the United States [J].
Bozic, Kevin J. ;
Kurtz, Steven ;
Lau, Edmund ;
Ong, Kevin ;
Chiu, Vanessa ;
Vail, Thomas P. ;
Rubash, Harry E. ;
Berry, Daniel J. .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2009, 91A (07) :1614-1620
[5]
An unusual lymphocytic perivascular infiltration in tissues around contemporary metal-on-metal joint replacements [J].
Davies, AP ;
Willert, HG ;
Campbell, PA ;
Learmonth, ID ;
Case, CP .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2005, 87A (01) :18-27
[6]
Metal Ion Measurement as a Diagnostic Tool to Identify Problems with Metal-on-Metal Hip Resurfacing [J].
de Smet, K. ;
de Haan, R. ;
Calistri, A. ;
Campbell, P. A. ;
Ebramzadeh, E. ;
Pattyn, C. ;
Gill, H. S. .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2008, 90A :202-208
[7]
ENGH CA, 2009, CLIN ORTHOP REL 0901
[8]
CT Outperforms Radiography for Determination of Acetabular Cup Version after THA [J].
Ghelman, Bernard ;
Kepler, Christopher K. ;
Lyman, Stephen ;
Della Valle, Alejandro Gonzalez .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2009, 467 (09) :2362-2370
[9]
Hip resurfacings revised for inflammatory pseudotumour have a poor outcome [J].
Grammatopoulos, G. ;
Pandit, H. ;
Kwon, Y. -M. ;
Gundle, R. ;
McLardy-Smith, P. ;
Beard, D. J. ;
Murray, D. W. ;
Gill, H. S. .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2009, 91B (08) :1019-1024
[10]
Loosening and osteolysis associated with metal-on-metal bearings: A local effect of metal hypersensitivity? [J].
Jacobs, Joshua J. ;
Hallab, Nadim J. .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2006, 88A (06) :1171-1172