In-line bipolar, steroid-eluting, high impedance, epimyocardial pacing lead

被引:2
作者
Karpawich, PP
Stokes, KB
Proctor, K
Schallhorn, R
McVenes, R
机构
[1] Wayne State Univ, Sch Med, Childrens Hosp Michigan, Cardiol Sect, Detroit, MI 48201 USA
[2] Medtron Inc, Minneapolis, MN USA
来源
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY | 1998年 / 21卷 / 03期
关键词
epimyocardial pacing; steroid; high impedance; bipolar;
D O I
10.1111/j.1540-8159.1998.tb00091.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Recent advances in electrode surface designs have eliminated traditional threshold differences between endo- and epicardial pacing leads. Since the epicardial approach offers the potential of direct left ventricular pacing and the transvenous approach may not be feasible or warranted in all instances, more advanced leads are being designed to optimize epicardial pacing capabilities. This study was conducted to evaluate a bipolar epimyocardial lead. Six immature canines (age 3 months) were instrumented. The lead (Medtronic model 10389) is a single-pass, "in-line" bipolar electrode with low current drain and high impedance, with an intramyocardial steroid-eluting cathode and nonsteroid epicardial anode. Twelve ventricular leads were implanted (two per animal) and the animals followed for 6 months with weekly analysis of pacing and sensing capabilities. Results at explant were compared with implant values and showed no significant differences between sensed R waves or in R wave slew rates in unipolar or bipolar modes. Explant lead impedances remained high in both modes: bipolar, 1658 +/- 331; and unipolar, 1327 +/- 308 Ohm (P < 0.05). Chronic voltage (V) threshold at 0.5 ms showed no significant change from implant values during the study: unipolar, 0.3 +/- 0.06 versus 1.0 +/- 0.8; and bipolar, 0.4 +/- 0.06 versus 1.6 +/- 1.2. Histologic review showed negligible fibrous reaction at the electrode-tissue interface. This study introduces a high impedance, low threshold, "in-line" bipolar pacing lead design capable of stable chronic pacing with implant facilitated by a single suture technique.
引用
收藏
页码:503 / 508
页数:6
相关论文
共 16 条
[1]   MECHANICAL PERFORMANCE AND EFFICIENCY OF LEFT VENTRICLE DURING VENTRICULAR STIMULATION [J].
BOERTH, RC ;
COVELL, JW .
AMERICAN JOURNAL OF PHYSIOLOGY, 1971, 221 (06) :1686-&
[2]  
Deanda Abe Jr., 1993, Journal of the American College of Cardiology, V21, p382A
[3]  
FIGA F, 1995, J AM COLL CARDIOL A, V53
[4]  
GOLDMAN NA, 1994, J AM COLL CARDIOL A, V349
[5]   PACEMAKER-INDUCED SUPERIOR VENA-CAVA SYNDROME - REPORT OF 4 CASES AND REVIEW OF THE LITERATURE [J].
GOUDEVENOS, JA ;
REID, PG ;
ADAMS, PC ;
HOLDEN, MP ;
WILLIAMS, DO .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1989, 12 (12) :1890-1895
[6]   STEROID-ELUTING EPICARDIAL LEADS IN PEDIATRICS - IMPROVED EPICARDIAL THRESHOLDS IN THE 1ST YEAR [J].
HAMILTON, R ;
GOW, R ;
BAHORIC, B ;
GRIFFITHS, J ;
FREEDOM, R ;
WILLIAMS, W .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1991, 14 (11) :2066-2072
[7]   EFFECTS OF ALTERED SITE OF ELECTRICAL ACTIVATION ON MYOCARDIAL PERFORMANCE DURING INOTROPIC STIMULATION [J].
HEYNDRICKX, GR ;
VILAINE, JP ;
KNIGHT, DR ;
VATNER, SF .
CIRCULATION, 1985, 71 (05) :1010-1016
[8]   STEROID-ELUTING EPICARDIAL PACING LEADS IN PEDIATRIC-PATIENTS - ENCOURAGING EARLY RESULTS [J].
JOHNS, JA ;
FISH, FA ;
BURGER, JD ;
HAMMON, JW .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1992, 20 (02) :395-401
[9]   SEPTAL HIS-PURKINJE VENTRICULAR PACING IN CANINES - A NEW ENDOCARDIAL ELECTRODE APPROACH [J].
KARPAWICH, PP ;
GATES, J ;
STOKES, KB .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1992, 15 (11) :2011-2015
[10]   FAILURE OF RATE RESPONSIVE VENTRICULAR PACING TO IMPROVE PHYSIOLOGICAL PERFORMANCE IN THE UNIVENTRICULAR HEART [J].
KARPAWICH, PP ;
PARIDON, SM ;
PINSKY, WW .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1991, 14 (11) :2058-2061