Evaluation of new resectoscope loop for transurethral resection of bladder tumors

被引:8
作者
Herr, HW [1 ]
Reuter, VE
机构
[1] Mem Sloan Kettering Canc Ctr, Urol Serv, New York, NY 10021 USA
[2] Mem Sloan Kettering Canc Ctr, Dept Surg, New York, NY 10021 USA
[3] Mem Sloan Kettering Canc Ctr, Dept Pathol, New York, NY 10021 USA
关键词
bladder neoplasms; cautery; urethra;
D O I
10.1016/S0022-5347(01)63249-X
中图分类号
R5 [内科学]; R69 [泌尿科学(泌尿生殖系疾病)];
学科分类号
1002 ; 100201 ;
摘要
Purpose: We evaluate a new resectoscope loop for transurethral resection of bladder tumors. Materials and Methods: Of 251 transurethral resections in 226 patients 111 were done with a conventional loop and 140 with the Olympus prototype model A2186 resectoscope loop. The quality of specimens provided for histological analysis was compared. Results: Tissue orientation was preserved and cautery artifact was reduced with the new loop compared to the standard resectoscope loop. Conclusions: Compared to a conventional resectoscope loop, the new loop resulted in a better tumor resection and helped facilitate histological evaluation of the tissue specimens.
引用
收藏
页码:2067 / 2068