Design and preliminary results of an ultrasound applicator for interstitial thermal coagulation

被引:56
作者
Lafon, C
Chapelon, JY
Prat, F
Gorry, F
Margonari, J
Theillere, Y
Cathignol, D
机构
[1] U281 INSERM, F-69424 Lyon, France
[2] Ctr Hosp Bicetre, F-94275 Le Kremlin Bicetre, France
关键词
ultrasound; interstitial; plane transducer; digestive tract; endoscopy;
D O I
10.1016/S0301-5629(97)00203-2
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
The extracorporeal high-intensity focused ultrasound (HIFU) techniques are not still transposable for tumors of the digestive tract because of their locations, This study was designed to evaluate the feasibility of interstitial applicators (3.8 mm O.D.) to comply with this therapeutic lack and to demonstrate the possibility of producing coagulation necrosis by a specially designed probe and with a short exposure time (20 s), The active surface of the applicators consists of plane water-cooled PZT transducers working at 10 or 5 MHz, They were evaluated in terms of acoustic power emitted as a function of the frequency, and applied electrical input (electroacoustic efficiency of 75% at their working frequency) and in vitro and in vivo pig liver tissue destruction, The in vitro and in vivo necroses depth from the applicator surface ranged from 8 to 20 mm, This showed the advantage of a nondivergent source: the pressure decay is only due to the tissue absorption in the Fresnel zone, The lesions dimensions are slightly dependent on perfusion: 8 +/- 2 mm deep in vitro for a 10.7-MHz transducer working at 14 W/cm(2) against 10 mm in vivo. Operating at 5 MHz makes it possible to increase the therapeutic heating depth, For example, at a similar close-to-transducer temperature, the 5-MHz applicator induced, at a depth of 10 mm, a temperature elevation of 27 degrees C against 15 degrees C for that working at 10 MHz. (C) 1998 World Federation for Ultrasound in Medicine & Biology.
引用
收藏
页码:113 / 122
页数:10
相关论文
共 18 条
  • [1] HIGH-INTENSITY FOCUSED ULTRASOUND FOR THE TREATMENT OF BENIGN PROSTATIC HYPERPLASIA - EARLY UNITED-STATES CLINICAL-EXPERIENCE
    BIHRLE, R
    FOSTER, RS
    SANGHVI, NT
    DONOHUE, JP
    HOOD, PJ
    [J]. JOURNAL OF UROLOGY, 1994, 151 (05) : 1271 - 1275
  • [2] EFFECTS OF PHYSICAL PARAMETERS ON HIGH-TEMPERATURE ULTRASOUND HYPERTHERMIA
    BILLARD, BE
    HYNYNEN, K
    ROEMER, RB
    [J]. ULTRASOUND IN MEDICINE AND BIOLOGY, 1990, 16 (04) : 409 - 420
  • [3] CHAPELON JY, 1992, CANCER RES, V52, P6353
  • [4] EFFECT OF BLOOD PERFUSION ON THE ABLATION OF LIVER PARENCHYMA WITH HIGH-INTENSITY FOCUSED ULTRASOUND
    CHEN, L
    TERHAAR, G
    HILL, CR
    DWORKIN, M
    CARNOCHAN, P
    YOUNG, H
    BENSTED, JPM
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 1993, 38 (11) : 1661 - 1673
  • [5] DAMIANOU CA, 1995 IEEE ULTR S, P1203
  • [6] Davidson F., 1991, OUTPUT MEASUREMENTS, P75
  • [7] FRY F J, 1978, Ultrasound in Medicine and Biology, V4, P337, DOI 10.1016/0301-5629(78)90022-4
  • [8] PRODUCTION OF FOCAL DESTRUCTIVE LESIONS IN THE CENTRAL NERVOUS SYSTEM WITH ULTRASOUND
    FRY, WJ
    MOSBERG, WH
    BARNARD, JW
    FRY, FJ
    [J]. JOURNAL OF NEUROSURGERY, 1954, 11 (05) : 471 - 478
  • [9] Gelet A, 1996, EUR UROL, V29, P174
  • [10] MRI-GUIDED NONINVASIVE ULTRASOUND SURGERY
    HYNYNEN, K
    DARKAZANLI, A
    UNGER, E
    SCHENCK, JF
    [J]. MEDICAL PHYSICS, 1993, 20 (01) : 107 - 115