Nonuniform strut distribution correlates with more neointimal hyperplasia after sirolimus-eluting stent implantation

被引:102
作者
Takebayashi, H [1 ]
Mintz, GS [1 ]
Carlier, SG [1 ]
Kobayashi, Y [1 ]
Fujii, K [1 ]
Yasuda, T [1 ]
Costa, RA [1 ]
Moussa, I [1 ]
Dangas, GD [1 ]
Mehran, R [1 ]
Lansky, AJ [1 ]
Kreps, E [1 ]
Collins, MB [1 ]
Colombo, A [1 ]
Stone, GW [1 ]
Leon, MB [1 ]
Moses, JW [1 ]
机构
[1] Cardiovasc Res Fdn, Lenox Hill Heart & Vasc Inst, New York, NY 10022 USA
关键词
restenosis; stents; ultrasonics;
D O I
10.1161/01.CIR.0000148371.53174.05
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background - Little is known about causes of intimal hyperplasia (IH) after sirolimus-eluting stent (SES) implantation. Methods and Results - Intravascular ultrasound was performed in 24 lesions with intra-SES restenosis and a comparison group of 25 nonrestenotic SESs. To assess stent strut distribution, the maximum interstrut angle was measured with a protractor centered on the stent, and the visible struts were counted and normalized for the number of stent cells. In SES restenosis patients, minimum lumen site was compared with image slices 2.5, 5.0, 7.5, and 10.0 mm proximal and distal to this site. The minimum lumen site had a smaller IVUS lumen area at follow-up (2.7 +/- 0.9 versus 6.2 +/- 1.9 mm(2); P < 0.01), larger maximum interstrut angle ( 135 +/- 39 degrees versus 72 +/- 23 degrees; P < 0.01), larger IH area (3.4 +/- 1.5 versus 0.6 +/- 1.1 mm(2); P < 0.01) and thickness (0.7 +/- 0.3 versus 0.1 +/- 0.2 mm; P < 0.01) at maximum interstrut angle, and fewer stent struts (4.9 +/- 1.0 versus 6.0 +/- 0.5; P < 0.01) even when normalized for the number of stent cells (0.78 +/- 0.15 versus 0.97 +/- 0.07; P < 0.01). Compared with nonrestenotic SES, the restenosis lesions also had a smaller minimal lumen area, larger IH area, thicker IH at maximum interstrut angle, fewer stent struts, and larger maximum interstrut angle. Multivariate analysis identified the number of visualized stent struts normalized for the number of stent cells and maximum interstrut angle as the only independent IVUS predictor of IH cross-sectional area ( P < 0.01 and P < 0.01), minimum lumen area ( P < 0.01 and P < 0.01), and IH thickness ( P < 0.01 and P < 0.01). Conclusions - The number and distribution of stent struts affect the amount of neointima after SES implantation.
引用
收藏
页码:3430 / 3434
页数:5
相关论文
共 10 条
[1]   Randomized study to assess the effectiveness of slow- and moderate-release polymer-based paclitaxel-eluting stents for coronary artery lesions [J].
Colombo, A ;
Drzewiecki, J ;
Banning, A ;
Grube, E ;
Hauptmann, K ;
Silber, S ;
Dudek, D ;
Fort, S ;
Schiele, F ;
Zmudka, K ;
Guagliumi, G ;
Russell, ME .
CIRCULATION, 2003, 108 (07) :788-794
[2]   Preliminary observations regarding angiographic pattern of restenosis after rapamycin-eluting stent implantation [J].
Colombo, A ;
Orlic, D ;
Stankovic, G ;
Corvaja, N ;
Spanos, V ;
Montorfano, M ;
Liistro, F ;
Carlino, M ;
Airoldi, F ;
Chieffo, A ;
Di Mario, C .
CIRCULATION, 2003, 107 (17) :2178-2180
[3]   Late incomplete lesion coverage following Cypher stent deployment for diffuse right coronary artery stenosis [J].
Halkin, A ;
Carlier, S ;
Leon, MB .
HEART, 2004, 90 (08) :e45
[4]   Physiological transport forces govern drug distribution for stent-based delivery [J].
Hwang, CW ;
Wu, D ;
Edelman, ER .
CIRCULATION, 2001, 104 (05) :600-605
[5]   Coronary restenosis after sirolimus-eluting stent implantation - Morphological description and mechanistic analysis from a consecutive series of cases [J].
Lemos, PA ;
Saia, F ;
Ligthart, JMR ;
Arampatzis, CA ;
Sianos, G ;
Tanabe, K ;
Hoye, A ;
Degertekin, M ;
Daemen, J ;
McFadden, E ;
Hofma, S ;
Smits, PC ;
de Feyter, P ;
van der Giessen, WJ ;
van Domburg, RT ;
Serruys, PW .
CIRCULATION, 2003, 108 (03) :257-260
[6]   A randomized comparison of a sirolimus-eluting stent with a standard stent for coronary revascularization. [J].
Morice, M ;
Serruys, PW ;
Sousa, JE ;
Fajadet, J ;
Hayashi, EB ;
Perin, M ;
Colombo, A ;
Schuler, G ;
Barragan, P ;
Guagliumi, G ;
Molnar, F ;
Falotico, R .
NEW ENGLAND JOURNAL OF MEDICINE, 2002, 346 (23) :1773-1780
[7]   Sirolimus-eluting stents versus standard stents in patients with stenosis in a native coronary artery [J].
Moses, JW ;
Leon, MB ;
Popma, JJ ;
Fitzgerald, PJ ;
Holmes, DR ;
O'Shaughnessy, C ;
Caputo, RP ;
Kereiakes, DJ ;
Williams, DO ;
Teirstein, PS ;
Jaeger, JL ;
Kuntz, RE .
NEW ENGLAND JOURNAL OF MEDICINE, 2003, 349 (14) :1315-1323
[8]  
Popma JJ., 2002, TEXT B INTERV CARDIO, P827
[9]   Stent fracture and restenosis in the drug-eluting stent era [J].
Sianos, G ;
Hofma, S ;
Ligthart, JMR ;
Saia, F ;
Hoye, A ;
Lemos, PA ;
Serruys, PW .
CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2004, 61 (01) :111-116
[10]   Two-year angiographic and intravascular ultrasound follow-up after implantation of sirolimus-eluting stents in human coronary arteries [J].
Sousa, JE ;
Costa, MA ;
Sousa, AGMR ;
Abizaid, AC ;
Seixas, AC ;
Abizaid, AS ;
Feres, F ;
Mattos, LA ;
Falotico, R ;
Jaeger, J ;
Popma, JJ ;
Serruys, PW .
CIRCULATION, 2003, 107 (03) :381-383