Validation of the Oscar 2 oscillometric 24-h ambulatory blood pressure monitor according to the British Hypertension Society protocol

被引:116
作者
Goodwin, James
Bilous, Mary
Winship, Susan
Finn, Paul
Jones, Stephen C.
机构
[1] James Cook Univ Hosp, Middlesbrough, Cleveland, England
[2] Univ Teesside, Middlesbrough, Cleveland, England
关键词
ambulatory devices; blood pressure; British Society of Hypertension protocol; Oscar; 2; oscillometric; SunTech;
D O I
10.1097/MBP.0b013e3280acab1b
中图分类号
R6 [外科学];
学科分类号
1002 ; 100210 ;
摘要
Objectives Accuracy of blood pressure measuring devices is of prime importance and should be validated before devices are used clinically. We carried out an independent evaluation of the Oscar 2 oscillometric ambulatory blood pressure monitor (SunTech Medical, Model 222) according to the British Hypertension Society (BHS) Protocol. Methods Validation of the Oscar 2 was carried out in accordance with Part 1 of the British Hypertension Society protocol. Having successfully completed the first three phases, in the static validation phase blood pressure measurements were made on 114 participants simultaneously by two observers blinded from each other's readings and those of the device, giving 255 data pairs for systolic and diastolic blood pressure. Readings were made using simultaneous same-arm measurement, each observer using a dual-head binaural stethoscope and a calibrated mercury sphygmomanometer. Data were recorded independently, to the nearest 2 mmHg and were checked by the supervisor who operated the device. Results The device mean and observer means (and standard deviations) were exceptionally close across a high range of pressures, with a mean difference for the better observer of 0 +/- 7 mmHg and - 1 +/- 6 mmHg for systolic and diastolic blood pressure, respectively. With 62% of all systolic blood pressure standard vs. device differences equal or less than 5 mmHg, and 70% of all similar differences for diastolic blood pressure equal to or less than 5 mmHg, the device was categorized as Grade A for systolic and diastolic blood pressure. Conclusion The Oscar 2 was graded A for both systolic and diastolic blood pressure by the British Hypertension Society protocol and can be recommended for clinical use in an adult population.
引用
收藏
页码:113 / 117
页数:5
相关论文
共 7 条
[1]   ABC of hypertension - The pathophysiology of hypertension [J].
Beevers, G ;
Lip, GYH ;
O'Brien, E .
BRITISH MEDICAL JOURNAL, 2001, 322 (7291) :912-916
[2]   The influence of the method of data analysis on the reported accuracy of automated blood pressure measuring devices [J].
Braam, RL ;
Thien, T .
JOURNAL OF HYPERTENSION, 2001, 19 (10) :1765-1767
[3]  
JAMIESON M, 1980, BLOOD PRESSURE MEASU
[4]   Validation of the OSCAR 2 oscillometric 24-hour ambulatory blood pressure monitor according to the International Protocol for the validation of blood pressure measuring devices [J].
Jones, SC ;
Bilous, M ;
Winship, S ;
Finn, P ;
Goodwin, J .
BLOOD PRESSURE MONITORING, 2004, 9 (04) :219-223
[5]   Working Group on Blood Pressure Monitoring of the European Society of Hypertension International Protocol for validation of blood pressure measuring devices in adults [J].
O'Brien, E ;
Pickering, T ;
Asmar, R ;
Myers, M ;
Parati, G ;
Staessen, J ;
Mengden, T ;
Imai, Y ;
Waeber, B ;
Palatini, P ;
Atkins, N ;
Gerin, W .
BLOOD PRESSURE MONITORING, 2002, 7 (01) :3-17
[6]  
O'Brien E., 1993, J HYPERTENSION S2, V11, P43, DOI DOI 10.1097/00004872-199306000-00013
[7]   RECOMMENDATIONS ON BLOOD-PRESSURE MEASUREMENT [J].
PETRIE, JC ;
OBRIEN, ET ;
LITTLER, WA ;
DESWIET, M .
BRITISH MEDICAL JOURNAL, 1986, 293 (6547) :611-615