Mitigation of susceptibility-induced signal loss in neuroimaging using localized shim coils

被引:66
作者
Hsu, JJ [1 ]
Glover, GH [1 ]
机构
[1] Stanford Univ, Sch Med, Lucas Magnet Resonance Spect & Imaging Ctr, Stanford, CA 94305 USA
关键词
localized active shimming; field homogeneity; brain MRI; fMRI; oral shim;
D O I
10.1002/mrm.20365
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Correction of magnetic field distortions is essential for obtaining accurate brain blood-oxygen-level-dependent functional magnetic resonance imaging (fMRI) activation maps. The present work introduces an active shimming method that utilizes the magnetic field generated by resistive shim coils placed in the mouth to locally homogenize the magnetic field in the inferior portion of the frontal lobe, where the field is most seriously distorted. The shimming field can be optimized in situ patient by patient for the region of interest of the scanner operator's choice. The method at 1.5 T is shown to be effective in reducing field inhomogeneity and in recovery of fMRI signal. For example, in a region of interest approximately of 149 cm(3), a coil of simple geometry can reduce the root mean square of the magnetic field by more than 50% and the recovered signal increases the extent of activation detected in a breath-holding fMRI experiment. (C) 2005 Wiley-Liss, Inc.
引用
收藏
页码:243 / 248
页数:6
相关论文
共 18 条
[1]   Self-navigated spiral fMRI: Interleaved versus single-shot [J].
Glover, GH ;
Lai, S .
MAGNETIC RESONANCE IN MEDICINE, 1998, 39 (03) :361-368
[2]   Improved combination of spiral-in/out images for BOLD fMRI [J].
Glover, GH ;
Thomason, ME .
MAGNETIC RESONANCE IN MEDICINE, 2004, 51 (04) :863-868
[3]  
HSU JJ, 2004, P 12 ANN M ISMRM KYO, P994
[4]  
HSU JJ, 2003, P 11 ANN M ISMRM TOR, P734
[5]   Functional magnetic resonance imaging of regional cerebral blood oxygenation changes during breath holding [J].
Kastrup, A ;
Li, TQ ;
Takahashi, A ;
Glover, GH ;
Moseley, ME .
STROKE, 1998, 29 (12) :2641-2645
[6]   DISCRIMINATION OF LARGE VENOUS VESSELS IN TIME-COURSE SPIRAL BLOOD-OXYGEN-LEVEL-DEPENDENT MAGNETIC-RESONANCE FUNCTIONAL NEUROIMAGING [J].
LEE, AT ;
GLOVER, GH ;
MEYER, CH .
MAGNETIC RESONANCE IN MEDICINE, 1995, 33 (06) :745-754
[7]   A COMPUTER-SIMULATION OF THE STATIC MAGNETIC-FIELD DISTRIBUTION IN THE HUMAN HEAD [J].
LI, SZ ;
WILLIAMS, GD ;
FRISK, TA ;
ARNOLD, BW ;
SMITH, MB .
MAGNETIC RESONANCE IN MEDICINE, 1995, 34 (02) :268-275
[8]   Three-dimensional mapping of the static magnetic field inside the human head [J].
Li, SZ ;
Dardzinski, BJ ;
Collins, CM ;
Yang, QX ;
Smith, MB .
MAGNETIC RESONANCE IN MEDICINE, 1996, 36 (05) :705-714
[9]   MAGNET FIELD PROFILING - ANALYSIS AND CORRECTING COIL DESIGN [J].
ROMEO, F ;
HOULT, DI .
MAGNETIC RESONANCE IN MEDICINE, 1984, 1 (01) :44-65
[10]  
ROOPCHANSINGH V, 2004, P 12 ANN M ISMRM KYO, P1650