A reference method for the analysis of aldosterone in blood by high-performance liquid chromatography atmospheric pressure chemical ionization tandem mass spectrometry

被引:52
作者
Fredline, VF [1 ]
Taylor, PJ [1 ]
Dodds, HM [1 ]
Johnson, AG [1 ]
机构
[1] UNIV QUEENSLAND,PRINCESS ALEXANDRA HOSP,DEPT MED,BRISBANE,QLD 4102,AUSTRALIA
关键词
D O I
10.1006/abio.1997.2340
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
A high-performance liquid chromatography-atmospheric pressure chemical ionization-tandem mass spectrometry (HPLC-APCI-MS/MS) reference method for the quantitation of aldosterone in serum and plasma has been developed. Samples were extracted with dichloromethane/diethyl ether, containing flumethasone as internal standard (IS). Chromatography was performed on a phenyl column using 50 mM ammonium formate (pH 7.1)/methanol (50/50, v/v) as mobile phase. Analysis was in negative-ionization mode by selected reaction monitoring (aldosterone m/z 359.2 --> 331.2; IS m/z 455.0 --> 379.0). The assay was linear over the range 15-500 pg/mL, with limits of detection and quantitation of 10 and 15 pg/mL, respectively. Imprecisions of the assay at 15, 20, 150, and 450 pg/mL were 18.5, 8.8, 10.6, and 9.5%, respectively. The accuracy of the method ranged from 93.1 to 98.9% with absolute recoveries between 84.0 and 91.3% (aldosterone) and 88.0 and 92.3% (IS). We present a case study of a patient admitted, with suspected primary hyperaldosteronism, on the basis of a high radioimmunoassay (RIA) aldosterone concentration. The results suggest that RIA was unreliable, causing unnecessary patient discomfort and a costly g-day hospital stay. The specific HPLC-API-MS/MS assay described offers the sensitivity and accuracy required to assess abnormal aldosterone production in hypertensive patients. (C) 1997 Academic Press.
引用
收藏
页码:308 / 313
页数:6
相关论文
共 26 条
[1]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[2]   On-line immunoaffinity chromatography high-performance liquid chromatography mass spectrometry for the determination of dexamethasone [J].
Creaser, CS ;
Feely, SJ ;
Houghton, E ;
Seymour, M ;
Teale, P .
ANALYTICAL COMMUNICATIONS, 1996, 33 (01) :5-8
[3]   OPTIMIZATION OF TOTAL URINARY ALDOSTERONE ESTIMATION - COMPARISON WITH OTHER LABORATORY METHODS FOR ASSESSMENT OF MINERALOCORTICOID STATUS [J].
CUNNINGHAM, SK ;
SEQUEIRA, SJ ;
CHAMBERS, J ;
HEFFERNAN, A ;
MCKENNA, TJ .
JOURNAL OF STEROID BIOCHEMISTRY AND MOLECULAR BIOLOGY, 1988, 31 (01) :125-130
[4]  
*DIAGN PROD CORP, 1990, PROT PERF COAT A COU
[5]   A high-performance liquid chromatography electrospray tandem mass spectrometry analysis of cortisol and metabolites in placental perfusate [J].
Dodds, HM ;
Taylor, PJ ;
Cannell, GR ;
Pond, SM .
ANALYTICAL BIOCHEMISTRY, 1997, 247 (02) :342-347
[6]   CORTISOL PRODUCTION-RATES MEASURED BY LIQUID-CHROMATOGRAPHY MASS-SPECTROMETRY [J].
ESTEBAN, NV ;
YERGEY, AL .
STEROIDS, 1990, 55 (04) :152-158
[7]   DETERMINATION OF SERUM CORTISOL BY THERMOSPRAY LIQUID-CHROMATOGRAPHY MASS-SPECTROMETRY - COMPARISON WITH GAS-CHROMATOGRAPHY MASS-SPECTROMETRY [J].
GASKELL, SJ ;
ROLLINS, K ;
SMITH, RW ;
PARKER, CE .
BIOMEDICAL AND ENVIRONMENTAL MASS SPECTROMETRY, 1987, 14 (12) :717-722
[8]   PRIMARY ALDOSTERONISM [J].
GORDON, RD .
JOURNAL OF ENDOCRINOLOGICAL INVESTIGATION, 1995, 18 (07) :495-511
[9]  
HANQUEZ C, 1958, J STEROID BIOCHEM, V31, P939
[10]   ENHANCED CHEMI-LUMINESCENT IMMUNOASSAY FOR ALDOSTERONE [J].
HUBL, W ;
THORPE, GHG ;
HOFMANN, F ;
MEISSNER, D ;
THIELE, HJ .
JOURNAL OF BIOLUMINESCENCE AND CHEMILUMINESCENCE, 1990, 5 (01) :49-52