A multicommuted flow system for sequential spectrophotometric determination of hydrosoluble vitamins in pharmaceutical preparations

被引:37
作者
Rocha, FRP
Fatibello, O
Reis, BF
机构
[1] Univ Sao Paulo, Ctr Energia Nucl Agr, BR-13400970 Piracicaba, SP, Brazil
[2] Univ Fed Sao Carlos, Dept Quim, BR-13560970 Sao Carlos, SP, Brazil
基金
巴西圣保罗研究基金会;
关键词
flow analysis; multicommutation; vitamin; sequential determination;
D O I
10.1016/S0039-9140(02)00477-0
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A multicommuted flow system is proposed for spectrophotometric determination of hydrosoluble vitamins (ascorbic acid, thiamine, riboflavine and pyridoxine) in pharmaceutical preparations. The flow manifold was designed with computer-controlled three-way solenoid valves for independent handling of sample and reagent solutions and a multi-channel spectrophotometer was employed for signal measurements. Periodic re-calibration as well as the standard addition method was implemented by using a single reference solution. Linear responses (r = 0.999) were obtained for 0.500-10.0 mg l(-1) ascorbic acid, 2.00-50.6 mgl(-1) thiamine, 5.00-50.0 mg l(-1) riboflavine and 0.500-8.00 mg l(-1) pyridoxine. Detection limits were estimated as 0.08 mg l(-1) (0.5 mumol l(-1)) ascorbic acid, 0.8 mg l(-1) (2 mumol l(-1)) thiamine, 0.2 mg l(-1) (0.5 mumol l(-1)) riboflavine and 0.1 mg l(-1) (0.9 mumol l(-1)) pyridoxine at 99.7% confidence level. A mean sampling rate of 60 determinations per hour was achieved and coefficients of variation of 1% (n = 20) were estimated for all species. The mean reagent consumption was 25-fold lower in relation to flow-based procedures with continuous reagent addition. Average recoveries between 95.6 and 100% were obtained for commercial pharmaceutical preparations. Results agreed with those obtained by reference methods at 95% confidence level. The flow system is suitable for application in quality control processes and in dissolution studies of vitamin tablets. (C) 2002 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:191 / 200
页数:10
相关论文
共 20 条
[1]  
Aniceto C, 1999, LAB ROBOTICS AUTOMAT, V11, P45, DOI 10.1002/(SICI)1098-2728(1999)11:1<45::AID-LRA6>3.0.CO
[2]  
2-7
[3]  
Aniceto C, 1999, QUIM NOVA, V22, P805
[4]   Spectrophotometric determination of vitamin B2 (riboflavin) in pharmaceutical formulations using flow injection analysis. [J].
Aniceto, C ;
Canaes, LD ;
Fatibello, O ;
Cavalheiro, CCS .
QUIMICA NOVA, 2000, 23 (05) :637-640
[5]  
*AOAC INT, 1995, OFF METH AN AOAC INT
[6]  
Fennema O.R., 1996, FOOD CHEM
[7]   High current-density anodic electro-dissolution in flow-injection systems for the determination of aluminium, copper and zinc in non-ferroalloys by flame atomic absorption spectrometry [J].
Giacomozzi, CA ;
Queiróz, RRU ;
Souza, IG ;
Neto, JAG .
JOURNAL OF AUTOMATED METHODS & MANAGEMENT IN CHEMISTRY, 1999, 21 (01) :17-22
[8]   Flow system based on a binary sampling process for automatic dilutions prior to flame atomic spectrometry [J].
Gine, MF ;
Packer, AP ;
Blanco, T ;
dosReis, BF .
ANALYTICA CHIMICA ACTA, 1996, 323 (1-3) :47-53
[9]   SPECTROPHOTOMETRIC FLOW-INJECTION DETERMINATION OF ASCORBIC-ACID BY GENERATION OF TRIIODIDE [J].
HERNANDEZMENDEZ, J ;
MATEOS, AA ;
PARRA, MJA ;
DEMARIA, CG .
ANALYTICA CHIMICA ACTA, 1986, 184 :243-250
[10]   BINARY SEARCH IN FLOW TITRATION EMPLOYING PHOTOMETRIC END-POINT DETECTION [J].
KORN, M ;
GOUVEIA, LFBP ;
DEOLIVEIRA, E ;
REIS, BF .
ANALYTICA CHIMICA ACTA, 1995, 313 (03) :177-184