Spectroscopic imaging with improved gradient modulated constant adiabaticity pulses on high-field clinical scanners

被引:76
作者
Andronesi, Ovidiu C. [1 ]
Ramadan, Saadallah [2 ]
Ratai, Eva-Maria [1 ]
Jennings, Dominique [1 ]
Mountford, Carolyn E. [2 ]
Sorensen, A. Gregory [1 ]
机构
[1] Massachusetts Gen Hosp, Dept Radiol, Martinos Ctr Biomed Imaging, Boston, MA 02114 USA
[2] Harvard Univ, Sch Med, Dept Radiol, Brigham & Womens Hosp,Ctr Clin Spect, Boston, MA 02115 USA
关键词
Magnetic resonance spectroscopic imaging (MRSI); Chemical shift displacement error (CSDE); Specific absorption rate (SAR); Gradient offset independent adiabatic (GOIA); Localized adiabatic selective refocusing (LASER); Brain; SPIN-ECHO PULSE; MAGNETIC-RESONANCE; HUMAN BRAIN; RADIOFREQUENCY PULSES; REFOCUSING PULSES; MR SPECTROSCOPY; FOCI PULSES; BROAD-BAND; INVERSION; NMR;
D O I
10.1016/j.jmr.2010.01.010
中图分类号
Q5 [生物化学];
学科分类号
070307 [化学生物学];
摘要
The purpose of this work was to design and implement constant adiabaticity gradient modulated pulses that have improved slice profiles and reduced artifacts for spectroscopic imaging on 3 T clinical scanners equipped with standard hardware. The newly proposed pulses were designed using the gradient offset independent adiabaticity (GOIA. Tannus and Garwood [13]) method using WURST modulation for RF and gradient waveforms. The GOIA-WURST pulses were compared with GOIA-HSn (GOIA based on nth-order hyperbolic secant) and FOCI (frequency offset corrected inversion) pulses of the same bandwidth and duration. Numerical simulations and experimental measurements in phantoms and healthy volunteers are presented. GOIA-WURST pulses provide improved slice profile that have less slice smearing for off-resonance frequencies compared to GOIA-HSn pulses. The peak RF amplitude of GOIA-WURST is much lower (40% less) than FOCI but slightly higher (14.9% more) to GOIA-HSn. The quality of spectra as shown by the analysis of lineshapes, eddy currents artifacts, subcutaneous lipid contamination and SNR is improved for GOIA-WURST. GOIA-WURST pulse tested in this work shows that reliable spectroscopic imaging could be obtained in routine clinical setup and might facilitate the use of clinical spectroscopy. (C) 2010 Elsevier Inc. All rights reserved.
引用
收藏
页码:283 / 293
页数:11
相关论文
共 40 条
[1]
ANDRONESI OC, 2009, P 17 ISMRM M HAW
[2]
[Anonymous], 1990, Principles of Nuclear Magnetic Resonance in One and Two Dimensions
[3]
ACUTE STROKE - EVALUATION WITH SERIAL PROTON MR SPECTROSCOPIC IMAGING [J].
BARKER, PB ;
GILLARD, JH ;
VANZIJL, PCM ;
SOHER, BJ ;
HANLEY, DF ;
AGILDERE, AM ;
OPPENHEIMER, SM ;
BRYAN, RN .
RADIOLOGY, 1994, 192 (03) :723-732
[4]
AZD2171, a pan-VEGF receptor tyrosine kinase inhibitor, normalizes tumor vasculature and alleviates edema in glioblastoma patients [J].
Batchelor, Tracy T. ;
Sorensen, A. Gregory ;
di Tomaso, Emmanuelle ;
Zhang, Wei-Ting ;
Duda, Dan G. ;
Cohen, Kenneth S. ;
Kozak, Kevin R. ;
Cahill, Daniel P. ;
Chen, Poe-Jou ;
Zhu, Mingwang ;
Ancukiewicz, Marek ;
Mrugala, Maciej M. ;
Plotkin, Scott ;
Drappatz, Jan ;
Louis, David N. ;
Ivy, Percy ;
Scadden, David T. ;
Benner, Thomas ;
Loeffler, Jay S. ;
Wen, Patrick Y. ;
Jain, Rakesh K. .
CANCER CELL, 2007, 11 (01) :83-95
[5]
BROAD-BAND AND ADIABATIC INVERSION OF A 2-LEVEL SYSTEM BY PHASE-MODULATED PULSES [J].
BAUM, J ;
TYCKO, R ;
PINES, A .
PHYSICAL REVIEW A, 1985, 32 (06) :3435-3447
[6]
SPATIAL LOCALIZATION IN NMR-SPECTROSCOPY INVIVO [J].
BOTTOMLEY, PA .
ANNALS OF THE NEW YORK ACADEMY OF SCIENCES, 1987, 508 :333-348
[7]
NMR CHEMICAL-SHIFT IMAGING IN 3 DIMENSIONS [J].
BROWN, TR ;
KINCAID, BM ;
UGURBIL, K .
PROCEEDINGS OF THE NATIONAL ACADEMY OF SCIENCES OF THE UNITED STATES OF AMERICA-BIOLOGICAL SCIENCES, 1982, 79 (11) :3523-3526
[8]
A SELECTIVE ADIABATIC SPIN-ECHO PULSE [J].
CONOLLY, S ;
NISHIMURA, D ;
MACOVSKI, A .
JOURNAL OF MAGNETIC RESONANCE, 1989, 83 (02) :324-334
[9]
VARIABLE-RATE SELECTIVE EXCITATION [J].
CONOLLY, S ;
NISHIMURA, D ;
MACOVSKI, A ;
GLOVER, G .
JOURNAL OF MAGNETIC RESONANCE, 1988, 78 (03) :440-458
[10]
A REDUCED POWER SELECTIVE ADIABATIC SPIN-ECHO PULSE SEQUENCE [J].
CONOLLY, S ;
GLOVER, G ;
NISHIMURA, D ;
MACOVSKI, A .
MAGNETIC RESONANCE IN MEDICINE, 1991, 18 (01) :28-38