Evaluation of a portable device based on peripheral arterial tone for unattended home sleep studies

被引:158
作者
Bar, A
Pillar, G
Dvir, I
Sheffy, J
Schnall, RP
Lavie, P
机构
[1] Technion Israel Inst Technol, Sleep Lab, Bruce Rappaport Fac Med, IL-32000 Haifa, Israel
[2] Itamar Med Ltd, Caesarea, Israel
关键词
ambulatory; automatic analysis; obstructive sleep apnea syndrome; peripheral arterial tone; respiratory disturbance index; sleep;
D O I
10.1378/chest.123.3.695
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Background: Diagnosis of obstructive sleep apnea syndrome (OSAS) by ambulatory systems is a growing practice in view of the large number of patients awaiting correct diagnosis. The Watch PAT100 (WP100) [Itamar Medical; Caesarea, Israel] is a portable device based on the peripheral arterial tone (PAT) signal, and is designed for unattended home sleep studies. Objectives: To evaluate the efficacy, reliability, and reproducibility of the WP100 device for the diagnosis of OSAS as compared to in-laboratory, standard polysomnographic-based manual scoring. Design and methods: One hundred two subjects (78 men; 69 patients with OSAS and 33 normal volunteers; mean+/-SD age, 41.4+/-15.2 years; body mass index, 26.8+/-5.5) underwent in-laboratory full polysomnography simultaneously with WP100 recording. Fourteen subjects also underwent two additional unattended home sleep studies with the WP100 alone. The polysomnography recordings were blindly scored for apnea/hypopnea according to the American Academy of Sleep Medicine criteria (1999), and the polysomnography respiratory disturbance index (RDI) [PSG-RDI] was calculated. The WP100 data were analyzed automatically for the PAT RDI (PRDI) by a proprietary algorithm that was previously developed on an independent group of subjects. Results: Across a wide range of RDI levels, the PRDI was highly correlated with the PSG-RDI (r=0.88, p<0.0001), with an area under the receiver operating characteristic curve of 0.82 and 0.87 for thresholds of 10 events per hour and 20 events per hour, respectively. The PRDI scores were also highly reproducible, showing high correlation between home and in-laboratory sleep studies (r=0.89, p<0.001). Conclusion: The WP100 may offer an accurate, robust, and reliable ambulatory method for the detection of OSAS, with minimal patient discomfort.
引用
收藏
页码:695 / 703
页数:9
相关论文
共 29 条
[1]   Nightly variability in the indices of sleep-disordered breathing in men being evaluated for impotence with consecutive night polysomnograms [J].
Chediak, AD ;
AcevedoCrespo, JC ;
Seiden, DJ ;
Kim, HH ;
Kiel, MH .
SLEEP, 1996, 19 (07) :589-592
[2]   Unreliability of automatic scoring of MESAM 4 in assessing patients with complicated obstructive sleep apnea syndrome [J].
Cirignotta, F ;
Mondini, S ;
Gerardi, R ;
Mostacci, B ;
Sancisi, E .
CHEST, 2001, 119 (05) :1387-1392
[3]   AMBULATORY MONITORING OF ARTERIAL OXYGEN-SATURATION [J].
DECKER, MJ ;
HOEKJE, PL ;
STROHL, KP .
CHEST, 1989, 95 (04) :717-722
[4]   Cost-effectiveness analysis of nocturnal oximetry as a method of screening for sleep apnea-hypopnea syndrome [J].
Epstein, LJE ;
Dorlac, GR .
CHEST, 1998, 113 (01) :97-103
[5]   Diagnostic accuracy of a portable recording device (MESAM IV) in suspected obstructive sleep apnoea [J].
Esnaola, S ;
Duran, J ;
InfanteRivard, C ;
Rubio, R ;
Fernandez, A .
EUROPEAN RESPIRATORY JOURNAL, 1996, 9 (12) :2597-2605
[6]   PORTABLE RECORDING IN THE ASSESSMENT OF OBSTRUCTIVE SLEEP-APNEA [J].
FERBER, R ;
MILLMAN, R ;
COPPOLA, M ;
FLEETHAM, J ;
MURRAY, CF ;
IBER, C ;
MCCALL, V ;
NINOMURCIA, G ;
PRESSMAN, M ;
SANDERS, M ;
STROHL, K ;
VOTTERI, B ;
WILLIAMS, A .
SLEEP, 1994, 17 (04) :378-392
[7]   Sleep-related breathing disorders in adults: Recommendations for syndrome definition and measurement techniques in clinical research [J].
Flemons, WW ;
Buysse, D ;
Redline, S ;
Pack, A ;
Strohl, K ;
Wheatley, J ;
Young, T ;
Douglas, N ;
Levy, P ;
McNicholas, W ;
Fleetham, J ;
White, D ;
Schmidt-Nowarra, W ;
Carley, D ;
Romaniuk, J .
SLEEP, 1999, 22 (05) :667-689
[8]  
Grote LB, 2001, SLEEP, V24, pA82
[9]   LOCAL REFLEX IN MICROCIRCULATION IN HUMAN CUTANEOUS TISSUE [J].
HENRIKSEN, O ;
SEJRSEN, P .
ACTA PHYSIOLOGICA SCANDINAVICA, 1976, 98 (02) :227-231
[10]   The medical cost of undiagnosed sleep apnea [J].
Kapur, V ;
Blough, DK ;
Sandblom, RE ;
Hert, R ;
de Maine, JB ;
Sullivan, SD ;
Psaty, BM .
SLEEP, 1999, 22 (06) :749-755