Field and laboratory tests of a high volume cascade impactor

被引:42
作者
Sillanpää, M
Hillamoa, R
Mäkelä, T
Pennanen, AS
Salonen, RO
机构
[1] Finnish Meteorol Inst, FIN-00880 Helsinki, Finland
[2] Natl Publ Hlth Inst, Dept Environm Hlth, FIN-70701 Kuopio, Finland
基金
芬兰科学院;
关键词
particulate matter; sampling methods; field comparisons laboratory calibration; collection efficiency;
D O I
10.1016/S0021-8502(02)00214-8
中图分类号
TQ [化学工业];
学科分类号
0817 ;
摘要
The performance of three different configurations of a recently developed high volume cascade impactor (HVCI) system was tested in both field and laboratory experiments. In the field, the HVCI was run simultaneously with a virtual impactor and a low pressure impactor using a 3- or 4-day sampling duration. The study was carried out in downtown Helsinki between June and September 2001. As the field test results were not consistent with the previous laboratory calibration results provided by the manufacturer, the collection efficiency of each field-tested HVCI stage and two additional stages was measured in the laboratory as a function of particle size, using monodisperse dioctyl sebacate test aerosol. The collection efficiency curve for stage 1, impacting particles on a greased aluminum plate, was steep. In stages 2-4, polyurethane foam was used as collection substrate, which made the collection efficiency curves less steep than that of stage 1. According to the present laboratory calibration and field experiments, the HVCI configuration that suits best for physicochemical and toxicological characterization of ambient air aerosol particles consists of an Andersen high volume PM10-inlet and of three HVCI stages with the 50% collection efficiency at 2.4, 0.9 and 0.2 mum in aerodynamic diameter, respectively. (C) 2003 Elsevier Science Ltd. All rights reserved.
引用
收藏
页码:485 / 500
页数:16
相关论文
共 22 条
[1]  
[Anonymous], 98 HLTH EFF I
[2]   MASS SIZE DISTRIBUTIONS OF TRAFFIC AEROSOLS AT VIENNA [J].
BERNER, A ;
LURZER, C .
JOURNAL OF PHYSICAL CHEMISTRY, 1980, 84 (16) :2079-2083
[3]   Development of a high volume cascade impactor for toxicological and chemical characterization studies [J].
Demokritou, P ;
Kavouras, IG ;
Ferguson, ST ;
Koutrakis, P .
AEROSOL SCIENCE AND TECHNOLOGY, 2002, 36 (09) :925-933
[4]   CALIBRATION OF THE QCM IMPACTOR FOR STRATOSPHERIC SAMPLING [J].
HERING, SV .
AEROSOL SCIENCE AND TECHNOLOGY, 1987, 7 (03) :257-274
[5]   ON THE PERFORMANCE OF THE BERNER LOW-PRESSURE IMPACTOR [J].
HILLAMO, RE ;
KAUPPINEN, EI .
AEROSOL SCIENCE AND TECHNOLOGY, 1991, 14 (01) :33-47
[6]  
Hinds W., 1999, Aerosol Technology: Properties, Behavior, and Measurement of Airborne Particles, DOI 10.1016/0021-8502(83)90049-6
[7]   Use of polyurethane foam as the impaction substrate/collection medium in conventional inertial impactors [J].
Kavouras, IG ;
Koutrakis, P .
AEROSOL SCIENCE AND TECHNOLOGY, 2001, 34 (01) :46-56
[8]  
Kavouras IG, 2000, INHAL TOXICOL, V12, P35, DOI 10.1080/08958378.2000.11463198
[9]   AEROSOL SAMPLING INLETS AND INHALABLE PARTICLES [J].
LIU, BYH ;
PUI, DYH .
ATMOSPHERIC ENVIRONMENT, 1981, 15 (04) :589-600
[10]   AEROSOL GENERATOR OF HIGH STABILITY [J].
LIU, BYH ;
LEE, KW .