Changes in stroke volume induced by passive leg raising in spontaneously breathing patients: comparison between echocardiography and Vigileo™/FloTrac™ device

被引:105
作者
Biais, Matthieu [1 ]
Vidil, Lionel [1 ]
Sarrabay, Philippe [1 ]
Cottenceau, Vincent [1 ]
Revel, Philippe [1 ]
Sztark, Francois [1 ]
机构
[1] CHU Bordeaux, Hop Pellegrin, Serv Anesthesie Reanimat 1, F-33076 Bordeaux, France
来源
CRITICAL CARE | 2009年 / 13卷 / 06期
关键词
CRITICALLY-ILL PATIENTS; MECHANICALLY VENTILATED PATIENTS; PREDICTING FLUID RESPONSIVENESS; OUTPUT MONITORING-SYSTEM; ARTERIAL PULSE PRESSURE; CARDIAC-OUTPUT; RESPIRATORY CHANGES; ICU PATIENTS; SURGERY; SHOCK;
D O I
10.1186/cc8195
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Introduction Passive leg raising (PLR) is a simple reversible maneuver that mimics rapid fluid loading and increases cardiac preload. The effects of this endogenous volume expansion on stroke volume enable the testing of fluid responsiveness with accuracy in spontaneously breathing patients. However, this maneuver requires the determination of stroke volume with a fast-response device, because the hemodynamic changes may be transient. The Vigileo (TM) monitor (Vigileo (TM); Flotrac (TM); Edwards Lifesciences, Irvine, CA, USA) analyzes systemic arterial pressure wave and allows continuous stroke volume monitoring. The aims of this study were (i) to compare changes in stroke volume induced by passive leg raising measured with the Vigileo (TM) device and with transthoracic echocardiography and (ii) to compare their ability to predict fluid responsiveness. Methods Thirty-four patients with spontaneous breathing activity and considered for volume expansion were included. Measurements of stroke volume were obtained with transthoracic echocardiography (SV-TTE) and with the Vigileo (TM) (SV-Flotrac) in a semi-recumbent position, during PLR and after volume expansion (500 ml saline). Patients were responders to volume expansion if SV-TTE increased >= 15%. Results Four patients were excluded. No patients received vasoactive drugs. Seven patients presented septic hypovolemia. PLR-induced changes in SV-TTE and in SV-Flotrac were correlated (r(2) = 0.56, P < 0.0001). An increase in SV-TTE >= 13% during PLR was predictive of response to volume expansion with a sensitivity of 100% and a specificity of 80%. An increase in SV-Flotrac >= 16% during PLR was predictive of response to volume expansion with a sensitivity of 85% and a specificity of 90%. There was no difference between the area under the ROC curve for PLR-induced changes in SV-TTE (AUC = 0.96 +/- 0.03) or SV-Flotrac (AUC = 0.92 +/- 0.05). Volume expansion-induced changes in SV-TTE correlated with volume expansion-induced changes in SV-Flotrac (r(2) = 0.77, P < 0.0001). In all patients, the highest plateau value of SV-TTE recorded during PLR was obtained within the first 90 s following leg elevation, whereas it was 120 s for SV-Flotrac. Conclusions PLR-induced changes in SV-Flotrac are able to predict the response to volume expansion in spontaneously breathing patients without vasoactive support.
引用
收藏
页数:8
相关论文
共 31 条
  • [1] Uncalibrated pulse contour-derived stroke volume variation predicts fluid responsiveness in mechanically ventilated patients undergoing liver transplantation
    Biais, M.
    Nouette-Gaulain, K.
    Cottenceau, V.
    Revel, P.
    Sztark, F.
    [J]. BRITISH JOURNAL OF ANAESTHESIA, 2008, 101 (06) : 761 - 768
  • [2] A Comparison of Stroke Volume Variation Measured by Vigileo™/FloTraC™ System and Aortic Doppler Echocardiography
    Biais, Matthieu
    Nouette-Gaulain, Karine
    Roullet, Stephanie
    Quinart, Alice
    Revel, Philippe
    Sztark, Francois
    [J]. ANESTHESIA AND ANALGESIA, 2009, 109 (02) : 466 - 469
  • [3] Evaluation of an uncalibrated arterial pulse contour cardiac output monitoring system in cirrhotic patients undergoing liver surgery
    Biancofiore, G.
    Critchley, L. A. H.
    Lee, A.
    Bindi, L.
    Bisa, M.
    Esposito, M.
    Meacci, L.
    Mozzo, R.
    DeSimone, P.
    Urbani, L.
    Filipponi, F.
    [J]. BRITISH JOURNAL OF ANAESTHESIA, 2009, 102 (01) : 47 - 54
  • [4] STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT
    BLAND, JM
    ALTMAN, DG
    [J]. LANCET, 1986, 1 (8476) : 307 - 310
  • [5] Changes in BP induced by passive leg raising predict response to fluid loading in critically ill patients
    Boulain, T
    Achard, JM
    Teboul, JL
    Richard, C
    Perrotin, D
    Ginies, G
    [J]. CHEST, 2002, 121 (04) : 1245 - 1252
  • [6] Hemodynamic effects of passive leg raising: an echocardiographic study in patients with shock
    Caille, Vincent
    Jabot, Julien
    Belliard, Guillaume
    Charron, Cyril
    Jardin, Francois
    Vieillard-Baron, Antoine
    [J]. INTENSIVE CARE MEDICINE, 2008, 34 (07) : 1239 - 1245
  • [7] Performance of a minimally invasive uncalibrated cardiac output monitoring system (Flotrac™/Vigileo™) in haemodynamically unstable patients
    Compton, F. D.
    Zukunft, B.
    Hoffmann, C.
    Zidek, W.
    Schaefer, J. -H.
    [J]. BRITISH JOURNAL OF ANAESTHESIA, 2008, 100 (04) : 451 - 456
  • [8] A meta-analysis of studies using bias and precision statistics to compare cardiac output measurement techniques
    Critchley, LAH
    Critchley, JAJH
    [J]. JOURNAL OF CLINICAL MONITORING AND COMPUTING, 1999, 15 (02) : 85 - 91
  • [9] Validation of a new arterial pulse contour-based cardiac output device
    de Waal, Eric E. C.
    Kalkman, Cor J.
    Rex, Steffen
    Buhre, Wolfgang F.
    [J]. CRITICAL CARE MEDICINE, 2007, 35 (08) : 1904 - 1909
  • [10] Respiratory changes in aortic blood velocity as an indicator of fluid responsiveness in ventilated patients with septic shock
    Feissel, M
    Michard, F
    Mangin, I
    Ruyer, O
    Faller, JP
    Teboul, JL
    [J]. CHEST, 2001, 119 (03) : 867 - 873