On-line separation and determination of trace amounts of manganese in biological samples by flow injection inductively coupled plasma atomic emission spectrometry

被引:18
作者
Zougagh, M [1 ]
de Torres, AG [1 ]
Pavón, JMC [1 ]
机构
[1] Univ Malaga, Fac Sci, Dept Analyt Chem, E-29071 Malaga, Spain
关键词
manganese; flow injection on-line preconcentration; DPTH-gel micro-column; Doehlert design; ICP-AES; biological samples;
D O I
10.1081/AL-120020147
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A procedure is developed for the determination of Mn in biological samples by FI-ICP-AES acid eluates after preconcentration of the samples. The preconcentration step is performed on a 3 cm x 3 mm column packed with silica gel functionalized with 1,5-bis(di-2-pyridyl)methylene thiocarbohydrazide (DPTH-gel) and placed in the injection valve of a simple flow manifold. The sorbed element is subsequently eluted with nitric acid. The acid eluate is analyzed by ICP-AES. Five variables (sample flow rate, eluent flow rate, eluent concentration, pH, and buffer concentration) were considered as factors in the optimization process. Interactions between analytical factors and their optimal levels were investigated using Response Surface Methodology (RSM) and based on sequential experimental Doehlert designs. The performance characteristics of the determination of manganese with DPTH-gel was: detection limits, 1.5 ng mL(-1), precision (RSD) 0.5% for 100 ng mL(-1), and enrichment factors, 26 using 60 s preconcentration time. The sampling frequency was 40 h(-1). The accuracy of the method was demonstrated by the analysis of certified reference biological samples. The results showed good agreement with the certified values.
引用
收藏
页码:1115 / 1130
页数:16
相关论文
共 27 条
[1]  
[Anonymous], PHARM EXPT DESIGN IN
[2]   USE OF EXPERIMENTAL-DESIGNS FOR ESTIMATION OF OPTIMUM ELECTROTHERMAL ATOMIC-ABSORPTION CONDITIONS FOR MOLYBDENUM [J].
ARAUJO, PW ;
GOMEZ, MJ ;
DEBENZO, ZA ;
CASTILLO, C .
CHEMOMETRICS AND INTELLIGENT LABORATORY SYSTEMS, 1992, 16 (03) :203-211
[3]   RECENT DEVELOPMENTS IN THE METHODS OF OPTIMUM AND RELATED EXPERIMENTAL-DESIGNS [J].
ATKINSON, AC .
INTERNATIONAL STATISTICAL REVIEW, 1988, 56 (02) :99-115
[4]  
Benkhedda I, 1999, J ANAL ATOM SPECTROM, V14, P957
[5]   SIMULTANEOUS-OPTIMIZATION OF PH AND ORGANIC MODIFIER CONTENT OF THE MOBILE PHASE FOR THE SEPARATION OF CHLOROPHENOLS USING A DOEHLERT DESIGN [J].
BOURGUIGNON, B ;
MARCENAC, F ;
KELLER, HR ;
DEAGUIAR, PF ;
MASSART, DL .
JOURNAL OF CHROMATOGRAPHY, 1993, 628 (02) :171-189
[6]  
Box G. E. P., 1960, TECHNOMETRICS, V2, P455, DOI [DOI 10.1080/00401706.1960.10489912, 10.2307/1266454]
[7]   ON THE EXPERIMENTAL ATTAINMENT OF OPTIMUM CONDITIONS [J].
BOX, GEP ;
WILSON, KB .
JOURNAL OF THE ROYAL STATISTICAL SOCIETY SERIES B-STATISTICAL METHODOLOGY, 1951, 13 (01) :1-45
[8]   MULTI-FACTOR EXPERIMENTAL-DESIGNS FOR EXPLORING RESPONSE SURFACES [J].
BOX, GEP ;
HUNTER, JS .
ANNALS OF MATHEMATICAL STATISTICS, 1957, 28 (01) :195-241
[9]   Automated on-line separation-preconcentration system for inductively coupled plasma atomic emission spectrometry and its application to mercury determination [J].
Rudner, PC ;
de Torres, AG ;
Pavon, JMC ;
Castellon, ER .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1998, 13 (04) :243-248
[10]  
Cela R, 1996, QUIM ANAL, V15, P5