Effectiveness of ammonia in reducing carbon-based polyatomic ion interferences in electrothermal vaporization collision cell inductively coupled plasma mass spectrometry

被引:24
作者
Ben-Younes, M
Gregoire, DC
Chakrabarti, CL
机构
[1] Geol Survey Canada, Ottawa, ON K1A 0E8, Canada
[2] Carleton Univ, Dept Chem, Ottawa, ON K1S 5B6, Canada
关键词
dynamic reaction cell; electrothermal vaporization; solution nebulization; ICP-MS; Cr; Mg; Si; polyatomic ions;
D O I
10.1016/S0584-8547(02)00149-0
中图分类号
O433 [光谱学];
学科分类号
0703 ; 070302 ;
摘要
The effectiveness of NH3 gas used in a Dynamic Reaction Cell (DRC) in reducing or eliminating carbon-based polyatomic ions in electrothermal vaporization (ETV) inductively coupled plasma mass spectrometry is reported. Ammonia gas was found effective in reducing interfering polyatomic ions on both Cr and Mg isotope ions. The use of NH3 as a reaction gas for the reduction of polyatomic ion interferences on Si was not fully successful. However, the LODs for Si-28(+) and Si-29(+) were greatly improved over conventional ICP-MS. Ammonia was found to be unsuitable as a reaction gas for the determination of Si-30(+). Results of signal-to background studies indicate that the experimental conditions required to optimize the DRC were close for solution nebulization and ETV for Cr only. For Mg and Si, different optimization conditions were required. (C) 2002 Elsevier Science B.V All rights reserved.
引用
收藏
页码:361 / 372
页数:12
相关论文
共 8 条
[1]  
Byrne JP, 1997, CAN J ANAL SCI SPECT, V42, P95
[2]   BACKGROUND SPECTRAL FEATURES IN ELECTROTHERMAL VAPORIZATION INDUCTIVELY-COUPLED PLASMA-MASS SPECTROMETRY - MOLECULAR-IONS RESULTING FROM THE USE OF CHEMICAL MODIFIERS [J].
GREGOIRE, DC ;
STURGEON, RE .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1993, 48 (11) :1347-1364
[3]  
Neubauer K, 1999, ATOM SPECTROSC, V20, P64
[4]   BACKGROUND SPECTRAL FEATURES IN INDUCTIVELY COUPLED PLASMA MASS-SPECTROMETRY [J].
TAN, SH ;
HORLICK, G .
APPLIED SPECTROSCOPY, 1986, 40 (04) :445-460
[5]  
TANNER S, 1998, 17 ANN SEM PUR WAT C, P183
[6]  
Tanner SD, 1999, ATOM SPECTROSC, V20, P45
[7]   A dynamic reaction cell for inductively coupled plasma mass spectrometry (ICP-DRC-MS). II. Reduction of interferences produced within the cell [J].
Tanner, SD ;
Baranov, VI .
JOURNAL OF THE AMERICAN SOCIETY FOR MASS SPECTROMETRY, 1999, 10 (11) :1083-1094
[8]   OXIDE, HYDROXIDE, AND DOUBLY CHARGED ANALYTE SPECIES IN INDUCTIVELY COUPLED PLASMA MASS-SPECTROMETRY [J].
VAUGHAN, MA ;
HORLICK, G .
APPLIED SPECTROSCOPY, 1986, 40 (04) :434-445