Measurement of the Stark shift of the Cs hyperfine splitting in an atomic fountain

被引:117
作者
Simon, E [1 ]
Laurent, P [1 ]
Clairon, A [1 ]
机构
[1] Observ Paris, BNM LPTF, F-75014 Paris, France
来源
PHYSICAL REVIEW A | 1998年 / 57卷 / 01期
关键词
D O I
10.1103/PhysRevA.57.436
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
We have used an atomic fountain frequency standard to perform a new measurement of the frequency shift induced by a static electric field E on the clock transition of the cesium 6(2)S(1/2) hyperfine splitting. The result in units of Hz/(V/m)(2) is delta nu(0) = -2.271(4) x 10(-10)E(2). The accuracy of this measurement represents a tenfold improvement over previous values. This result is of importance for testing ab initio theoretical calculations of the cesium atomic structure as well as semiempirical models used in parity violation experiments. It will also allow a better evaluation of the frequency shift induced by the blackbody radiation in cesium frequency standards. [S1050-2947(98)02901-1].
引用
收藏
页码:436 / 439
页数:4
相关论文
共 19 条
[1]   HYPERFINE STRUCTURE STARK EFFECT .I. THEORY [J].
ANGEL, JRP ;
SANDARS, PGH .
PROCEEDINGS OF THE ROYAL SOCIETY OF LONDON SERIES A-MATHEMATICAL AND PHYSICAL SCIENCES, 1968, 305 (1480) :125-&
[2]   Experimental verification of the shift of the cesium hyperfine transition frequency due to blackbody radiation [J].
Bauch, A ;
Schroder, R .
PHYSICAL REVIEW LETTERS, 1997, 78 (04) :622-625
[3]   ATOMIC-BEAM RESONANCE MEASUREMENT OF DIFFERENTIAL POLARIZABILITY BETWEEN ZEEMAN SUBSTATES IN GROUND STATE OF CESIUM ATOM [J].
CARRICO, JP ;
ADLER, A ;
BAKER, MR ;
LEGOWSKI, S ;
LIPWORTH, E ;
SANDARS, PGH ;
STEIN, TS ;
WEISSKOPF, MC .
PHYSICAL REVIEW, 1968, 170 (01) :64-+
[4]   RAMSEY RESONANCE IN A ZACHARIAS FOUNTAIN [J].
CLAIRON, A ;
SALOMON, C ;
GUELLATI, S ;
PHILLIPS, WD .
EUROPHYSICS LETTERS, 1991, 16 (02) :165-170
[5]   A CESIUM FOUNTAIN FREQUENCY STANDARD - PRELIMINARY-RESULTS [J].
CLAIRON, A ;
LAURENT, P ;
SANTARELLI, G ;
GHEZALI, S ;
LEA, SN ;
BAHOURA, M .
IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 1995, 44 (02) :128-131
[6]  
Clairon A., 1996, Proceedings of the Fifth Symposium on Frequency Standards and Metrology, P49
[7]   STARK EFFECT OF HYPERFINE STRUCTURE OF CESIUM-133 [J].
FEICHTNER, JD ;
HOOVER, ME ;
MIZUSHIMA, M .
PHYSICAL REVIEW, 1965, 137 (3A) :A702-+
[8]   An experimental study of the spin-exchange frequency shift in a laser-cooled cesium fountain frequency standard [J].
Ghezali, S ;
Laurent, P ;
Lea, SN ;
Clairon, A .
EUROPHYSICS LETTERS, 1996, 36 (01) :25-30
[9]   LASER-COOLED CS FREQUENCY STANDARD AND A MEASUREMENT OF THE FREQUENCY-SHIFT DUE TO ULTRACOLD COLLISIONS [J].
GIBBLE, K ;
CHU, S .
PHYSICAL REVIEW LETTERS, 1993, 70 (12) :1771-1774
[10]   FUTURE SLOW-ATOM FREQUENCY STANDARDS [J].
GIBBLE, K ;
CHU, S .
METROLOGIA, 1992, 29 (02) :201-212