Validation of a vibrating-wire viscometer:: Measurements in the range of 0.5 to 135 mPa•s

被引:42
作者
Caetano, FJP
Fareleira, JMNA
Oliveira, CMBP
Wakeham, WA
机构
[1] Univ Tecn Lisboa, Inst Super Tecn, Ctr Quim Estrutural, P-1049001 Lisbon, Portugal
[2] Univ Alberta, R Escola Politecn, P-1269001 Lisbon, Portugal
[3] Univ Southampton, Southampton SO17 1BJ, Hants, England
关键词
D O I
10.1021/je049733f
中图分类号
O414.1 [热力学];
学科分类号
摘要
The present article describes a novel vibrating-wire viscometer that has been shown to be able to measure viscosities up to 135 mPa(.)s, after calibration against water at 20 degreesC, to provide direct traceability to the primary reference for viscosity. For the purpose of validating the instrument, measurements of the viscosity of some selected fluids, including standard reference liquids with viscosity on the order of 100 mPa(.)s at 20 degreesC and 2,2,4-trimethylpentane with a viscosity of about 0.5 mPa(.)s at the same temperature, were performed. The results obtained show that the instrument is capable of performing viscosity measurements with an estimated overall uncertainty better than +/-0.8% over the range of (0.5 to 135) mPa(.)s.
引用
收藏
页码:201 / 205
页数:5
相关论文
共 16 条
[1]   PRECISE VISCOSITY MEASUREMENTS OF NEWTONIAN LIQUIDS WITH V LESS-THAN 1 MM2/S FOR THE SELECTION OF SUITABLE STANDARDS [J].
BAUER, H ;
MEERLENDER, G .
RHEOLOGICA ACTA, 1984, 23 (05) :514-521
[2]  
BETTIN H, 1990, PTB-MITT, V100, P195
[3]   Viscosity of di-isodecylphthalate: A potential standard of moderate viscosity [J].
Caetano, FJP ;
Fareleira, JMNA ;
Oliveira, CMBP ;
Wakeham, WA .
INTERNATIONAL JOURNAL OF THERMOPHYSICS, 2004, 25 (05) :1311-1322
[4]   Viscosity measurements of liquid toluene at low temperatures using a dual vibrating-wire technique [J].
Caetano, FJP ;
da Mata, JLC ;
Fareleira, JMNA ;
Oliveira, CMBP ;
Wakeham, WA .
INTERNATIONAL JOURNAL OF THERMOPHYSICS, 2004, 25 (01) :1-11
[5]  
CAETANO FJP, UNPUB
[6]   ADDED MASS AND DAMPING OF A VIBRATING ROD IN CONFINED VISCOUS FLUIDS [J].
CHEN, SS ;
WAMBSGANSS, MW ;
JENDRZEJCZYK, JA .
JOURNAL OF APPLIED MECHANICS-TRANSACTIONS OF THE ASME, 1976, 43 (02) :325-329
[7]  
Oliveira C.M.B.P., 1991, THESIS IMPERIAL COLL
[8]   Electromechanical model for vibrating-wire instruments [J].
Padua, AAH ;
Fareleira, JMNA ;
Calado, JCG ;
Wakeham, WA .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1998, 69 (06) :2392-2399
[9]   Validation of an accurate vibrating-wire densimeter: Density and viscosity of liquids over wide ranges of temperature and pressure [J].
Padua, AAH ;
Fareleira, JMNA ;
Calado, JCG ;
Wakeham, WA .
INTERNATIONAL JOURNAL OF THERMOPHYSICS, 1996, 17 (04) :781-802
[10]   Density and viscosity measurements of 2,2,4-trimethylpentane (isooctane) from 198 K to 348 K and up to 100 MPa [J].
Padua, AAH ;
Fareleira, JMNA ;
Calado, JCG ;
Wakeham, WA .
JOURNAL OF CHEMICAL AND ENGINEERING DATA, 1996, 41 (06) :1488-1494