Magnetically suspended centrifugal blood pump with a radial magnetic driver

被引:30
作者
Hoshi, H
Katakoa, K
Ohuchi, K
Asama, J
Shinshi, T
Shimokohbe, A
Takatani, S
机构
[1] Tokyo Med & Dent Univ, Dept Artificial Organs, Inst Biomat & Bioengn, Chiyoda Ku, Tokyo 1010062, Japan
[2] Tokyo Inst Technol, Inst Precis Control Engn, Kanagawa, Japan
关键词
D O I
10.1097/01.mat.0000151148.72891.55
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
A new magnetic bearing has been designed to achieve a low electronic power requirement and high stiffness. The magnetic bearing consists of 1) radial passive forces between the permanent magnet ring mounted inside the impeller rotor and the electromagnet core materials in the pump casing and 2) radial active forces generated by the electromagnets using the two gap sensor signals. The magnetic bearing was assembled into a centrifugal rotary blood pump (CRBP) driven with a radial, magnetic coupled driver. The impeller vane shape was designed based upon the computational fluid dynamic simulation. The diameter and height of the CRBP were 75 mm and 50 mm, respectively. The magnetic bearing system required the power of 1.0-1.4 W. The radial impeller movement was controlled to within +/-10 mum. High stiffness in the noncontrolled axes, Z, Phi, and theta, was obtained by the passive magnetic forces. The pump flow of 5 L/min against 100 mm Hg head pressure was obtained at 1,800 rpm with the electrical to hydraulic efficiency being greater than 15%. The Normalized Index of Hemolysis (NIH) of the magnetic bearing CRBP was one fifth of the BioPump BP-80 and one half of the NIKKISO HPM-15 after 4 hours. The newly designed magnetic bearing with two degrees of freedom control in combination with optimized impeller vane was successful in achieving an excellent hemolytic performance in comparison with the clinical centrifugal blood pumps.
引用
收藏
页码:60 / 64
页数:5
相关论文
共 15 条
[1]  
ASAMA J, 2003, P 7 INT S MAGN SUSP, P117
[2]   HeartMate III: Pump design for a centrifugal LVAD with a magnetically levitated rotor [J].
Bourque, K ;
Gernes, DB ;
Loree, HM ;
Richardson, JS ;
Poirier, VL ;
Barletta, N ;
Fleischli, A ;
Foiera, G ;
Gempp, TM ;
Schoeb, R ;
Litwak, KN ;
Akimoto, T ;
Watach, MJ ;
Litwak, P .
ASAIO JOURNAL, 2001, 47 (04) :401-405
[3]   Impeller behavior and displacement of the VentrAssist implantable rotary blood pump [J].
Chung, MKH ;
Zhang, N ;
Tansley, TD ;
Woodard, JC .
ARTIFICIAL ORGANS, 2004, 28 (03) :287-297
[4]   Preclinical readiness testing of the Arrow International CorAide left ventricular assist system [J].
Doi, K ;
Golding, LAR ;
Massiello, AL ;
Kopcak, MW ;
Gerhart, RL ;
Schenk, S ;
Inoue, M ;
Ootaki, Y ;
Fukamachi, K .
ANNALS OF THORACIC SURGERY, 2004, 77 (06) :2103-2110
[5]   Worldwide experience with the MicroMed DeBakey ventricular assist device® as a bridge to transplantation [J].
Goldstein, DJ .
CIRCULATION, 2003, 108 (10) :272-277
[6]   Magnetically suspended centrifugal blood pump with an axially levitated motor [J].
Masuzawa, T ;
Ezoe, S ;
Kato, T ;
Okada, Y .
ARTIFICIAL ORGANS, 2003, 27 (07) :631-638
[7]   Terumo implantable left ventricular assist system: Results of long-term animal study [J].
Nojiir, C ;
Kijima, T ;
Maekawa, J ;
Horiuchi, K ;
Kido, T ;
Sugiyama, T ;
Mori, T ;
Sugiura, N ;
Asada, T ;
Ozaki, T ;
Suzuki, M ;
Akamatsu, T ;
Akutsu, T .
ASAIO JOURNAL, 2000, 46 (01) :117-122
[8]   Long-term use of a left ventricular assist device for end-stage heart failure [J].
Rose, EA ;
Gelijns, AC ;
Moskowitz, AJ ;
Heitjan, DF ;
Stevenson, LW ;
Dembitsky, W ;
Long, JW ;
Ascheim, DD ;
Tierney, AR ;
Levitan, RG ;
Watson, JT ;
Meier, P .
NEW ENGLAND JOURNAL OF MEDICINE, 2001, 345 (20) :1435-1443
[9]  
SHINSHI T, 2003, P IIP ISOS JOINT MIP, P377