A microfluidic device based on gravity and electric force driving for flow cytometry and fluorescence activated cell sorting

被引:118
作者
Yao, B
Luo, GA [1 ]
Feng, X
Wang, W
Chen, LX
Wang, YM
机构
[1] Tsinghua Univ, Dept Chem, Beijing 100084, Peoples R China
[2] Tsinghua Univ, Dept Engn Mech, Beijing 100084, Peoples R China
关键词
D O I
10.1039/b408422e
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
A novel method based on gravity and electric force driving of cells was developed for flow cytometry and fluorescence activated cell sorting in a microfluidic chip system. In the experiments cells flowed spontaneously under their own gravity in a upright microchip, passed through the detection region and then entered into the sorting electric field one by one at an average velocity of 0.55 mm s(-1) and were fluorescence activated cell sorted (FACS) by a switch-off activation program. In order to study the dynamical and kinematic characteristics of single cells in gravity and electric field of microchannels a physical and numerical module based on Newton's Law of motion was established and optimized. Hydroxylpropylmethyl cellulose (HPMC) was used to minimize cell assembling, sedimentation and adsorption to microchannels. This system was applied to estimate the necrotic and apoptotic effects of ultraviolet (UV) light on HeLa cells by exposing them to UV radiation for 10, 20 or 40 min and the results showed that UV radiation induced membrane damage contributed to the apoptosis and necrosis of HeLa cells.
引用
收藏
页码:603 / 607
页数:5
相关论文
共 46 条
[1]   Ultraviolet light induces apoptosis via direct activation of CD95 (Fas/APO-1) independently of its ligand CD95L [J].
Aragane, Y ;
Kulms, D ;
Metze, D ;
Wilkes, G ;
Pöppelmann, B ;
Luger, TA ;
Schwarz, T .
JOURNAL OF CELL BIOLOGY, 1998, 140 (01) :171-182
[2]  
Bergkvist J, 2002, PROTEOMICS, V2, P422, DOI 10.1002/1615-9861(200204)2:4<422::AID-PROT422>3.0.CO
[3]  
2-1
[4]  
Bink K, 2001, HISTOCHEM CELL BIOL, V115, P293
[5]   Passively driven integrated microfluidic system for separation of motile sperm [J].
Cho, BS ;
Schuster, TG ;
Zhu, XY ;
Chang, D ;
Smith, GD ;
Takayama, S .
ANALYTICAL CHEMISTRY, 2003, 75 (07) :1671-1675
[6]  
Chung S, 2003, MICROSYST TECHNOL, V9, P525, DOI [10.1007/s00542-003-0302-2, 10.1007/S00542-003-0302-2]
[7]   Optical particle detection integrated in a dielectrophoretic lab-on-a-chip [J].
Cui, L ;
Zhang, T ;
Morgan, H .
JOURNAL OF MICROMECHANICS AND MICROENGINEERING, 2002, 12 (01) :7-12
[8]  
Cui L, 2001, ELECTROPHORESIS, V22, P3893, DOI 10.1002/1522-2683(200110)22:18<3893::AID-ELPS3893>3.0.CO
[9]  
2-2
[10]   An integrated microfluidic system for reaction, high-sensitivity detection, and sorting of fluorescent cells and particles [J].
Dittrich, PS ;
Schwille, P .
ANALYTICAL CHEMISTRY, 2003, 75 (21) :5767-5774