Quality control of mechanical ventilation at the patient's home

被引:13
作者
Farré, R
Giró, E
Casolivé, V
Navajas, D
Escarrabill, J
机构
[1] Univ Barcelona, Inst Invest Biomed August Pi Sunyer, Fac Med, Unitat Biofis & Bioengn, E-08036 Barcelona, Spain
[2] Univ Barcelona, Bellvitge Hosp, UFISS Resp, Serv Pneumol, Barcelona, Spain
关键词
home mechanical ventilation; chronic respiratory failure; ventilator performance;
D O I
10.1007/s00134-002-1593-y
中图分类号
R4 [临床医学];
学科分类号
1002 [临床医学]; 100602 [中西医结合临床];
摘要
Objective: During home mechanical ventilation the prescribed settings are applied without permanent supervision of health professionals. After a long-time period of unattended operation at home the ventilator may not apply the ventilation parameters prescribed. This quality control study of home mechanical ventilation assessed whether tidal volume (V-T), frequency (f), and minute ventilation (V'(E)) actually applied by the ventilator coincide with the values set on the ventilator control panel and with those prescribed. Measurements: Actual V-T, f, and V'(E) applied by the ventilator in 30 patients on nocturnal HMV were measured at the patients' homes. The patients were subjected to volume targeted assist ventilation through nasal mask (n = 28) or tracheostomy ( n = 2). The values of V-T, f, and V'(E) set at the ventilator were recorded. The actual and set V-T, f, and V'(E) values were compared with those prescribed. Results: Considerable differences were found between actual, set and prescribed V-T, f, and V'(E). Actual VE was significantly lower than VE set: mean difference was 0.82 l/min, with considerable individual differences. Differences between actual and prescribed VE were caused both by a poor performance of the ventilator and by a discrepancy between the values prescribed and those set at the ventilator control panel. Conclusions: Regularly assessing the actual performance of ventilators at the patient's home is a quality control procedure useful for detecting malfunctions which could improve compliance and outcome of home mechanical ventilation.
引用
收藏
页码:484 / 486
页数:3
相关论文
共 6 条
[1]
Long-term home care programmes may reduce hospital admissions in COPD with chronic hypercapnia [J].
Clini, E ;
Vitacca, M ;
Foglio, K ;
Simoni, P ;
Ambrosino, N .
EUROPEAN RESPIRATORY JOURNAL, 1996, 9 (08) :1605-1610
[2]
Noninvasive mechanical ventilation at home - Building upon the tradition [J].
Goldberg, AI .
CHEST, 2002, 121 (02) :321-324
[3]
A fatal complication of noninvasive ventilation. [J].
Lechtzin, N ;
Weiner, CM ;
Clawson, L .
NEW ENGLAND JOURNAL OF MEDICINE, 2001, 344 (07) :533-533
[4]
Inaccuracy of tidal volume delivered by home mechanical ventilators [J].
Lofaso, F ;
Fodil, R ;
Lorino, H ;
Leroux, K ;
Quintel, A ;
Leroy, A ;
Harf, A .
EUROPEAN RESPIRATORY JOURNAL, 2000, 15 (02) :338-341
[5]
RAMOS PD, 2000, ARCH BRONCONEUMOL, V36, P545
[6]
Frequency, causes, and outcome of home ventilator failure [J].
Srinivasan, S ;
Doty, SM ;
White, TR ;
Segura, VH ;
Jansen, MT ;
Ward, SLD ;
Keens, TG .
CHEST, 1998, 114 (05) :1363-1367