Biomechanical evaluation of the dynamic hip screw with two- and four-hole side plates

被引:89
作者
McLoughlin, SW
Wheeler, DL
Rider, J
Bolhofner, B
机构
[1] Oregon Hlth Sci Univ, Dept Orthopaed & Rehabil, Portland, OR 97201 USA
[2] All Florida Orthopaed Associates, St Petersburg, FL USA
关键词
intertrochanteric hip fracture; biomechanics; dynamic hip screw;
D O I
10.1097/00005131-200006000-00002
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Objectives: To determine the biomechanical strength and stiffness of a dynamic hip screw (DHS; Synthes USA, Paoli, PA, U.S.A.) with a two-hole side-plate as compared with a four-hole side-plate design for the reconstruction of unstable three-part intertrochanteric fractures. Design: Eight matched pairs of embalmed human femurs were tested in two modes: (a) 2,000 cycles of simulated physiologic loading; (b) test to failure. Setting: Laboratory. Simulated single leg stance using a simulated pelvic loading mechanism with abductor loading. Strain and displacement sensors were used to measure fragment shear and distraction and surface strain in the proximal side plate. Specimens: Eight pairs of skeletonized embalmed cadaveric specimens were selected on the basis of femoral neck angle and absence of old fracture, anatomic anomaly, or pathology. Intervention: The specimens were divided into two groups: ia) left femurs received the two-hole side-plate design; (b) right femurs received the four-hole side-plate design. All fractures were reconstructed by the same surgeon using the manufacturer's instructions. Main Outcome Measurements: Implant placement was verified by radiographic measurement of tip-to-apex distance. In cyclic testing, the amount of femoral neck fragment migration in both distraction and shear was quantified. Strain magnitude in the side plate was measured in bath cyclic and failure testing. The peak load withstood by the reconstruction was quantified in the failure test. Results and Conclusions: Peak load in the failure test was not found to be statistically different between the two-hole and four-hole designs. In cyclic testing, the two-hole configuration exhibited statistically smaller fragment migration in both shear and distraction than the four-hole design (p < 0.05). The strain magnitude in the side plate was not statistically different in the cyclic or failure tests. The femurs with a greater neck angle failed by crushing of the bone in the neck. The femurs with a lesser neck angle failed due to bending of the hardware. The results of this investigation revealed that the two-hole DHS is biomechanically as stable as the four-hole DHS in cyclic and failure loads under the conditions tested. These results, in concert with clinical experience, can be used to support the use of the two-hole DHS for the reconstruction of intertrochanteric fractures without a diaphyseal extension.
引用
收藏
页码:318 / 323
页数:6
相关论文
共 19 条
[1]
Periprosthetic bone remodelling after cementless total hip replacement - A prospective comparison of two different implant designs [J].
Ang, KC ;
De, SD ;
Goh, JCH ;
Low, SL ;
Bose, K .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1997, 79B (04) :675-679
[2]
THE VALUE OF THE TIP-APEX DISTANCE IN PREDICTING FAILURE OF FIXATION OF PERITROCHANTERIC FRACTURES OF THE HIP [J].
BAUMGAERTNER, MR ;
CURTIN, SL ;
LINDSKOG, DM ;
KEGGI, JM .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1995, 77A (07) :1058-1064
[3]
CUMMINGS SR, 1990, CLIN ORTHOP RELAT R, P163
[4]
A BIOMECHANICAL EVALUATION OF AN INTRAMEDULLARY FIXATION DEVICE FOR INTERTROCHANTERIC FRACTURES [J].
FLAHIFF, CM ;
NELSON, CL ;
GRUENWALD, JM ;
HOLLIS, JM .
JOURNAL OF TRAUMA-INJURY INFECTION AND CRITICAL CARE, 1993, 35 (01) :23-27
[5]
HOW TO MINIMIZE FAILURES OF FIXATION OF UNSTABLE INTERTROCHANTERIC FRACTURES [J].
GUNDLE, R ;
GARGAN, MF ;
SIMPSON, AHRW .
INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 1995, 26 (09) :611-614
[6]
Fixation of intracapsular femoral neck fractures with a one-hole plate dynamic hip screw [J].
HeyseMoore, GH .
INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 1996, 27 (03) :181-183
[7]
Weight bearing after hip fracture: A prospective series of 596 geriatric hip fracture patients [J].
Koval, KJ ;
Friend, KD ;
Aharonoff, GB ;
Zuckerman, JD .
JOURNAL OF ORTHOPAEDIC TRAUMA, 1996, 10 (08) :526-530
[8]
Koval KJ, 1996, CLIN ORTHOP RELAT R, P166
[9]
LEVY RN, 1992, SKELETAL TRAUMA, V2, P1443
[10]
MICHELSON JD, 1995, CLIN ORTHOP RELAT R, P129