Comparison of transthoracic three dimensional echocardiography with magnetic resonance imaging in the assessment of right ventricular volume and mass

被引:125
作者
Vogel, M [1 ]
Gutberlet, M [1 ]
Dittrich, S [1 ]
Hosten, N [1 ]
Lange, PE [1 ]
机构
[1] HUMBOLDT UNIV BERLIN, UNIV HOSP RUDOLF VIRCHOW, DEPT RADIOL, BERLIN, GERMANY
关键词
right ventricular volume; right ventricular mass; three dimensional echocardiography; magnetic resonance imaging;
D O I
10.1136/hrt.78.2.127
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Objective-Assessment of right ventricular volume and mass with three dimensional echocardiography and comparison with magnetic resonance imaging. Methods-Measurements of right ventricular volumes performed on three dimensional datasets acquired by transthoracic echocardiography were compared to those obtained from magnetic resonance imaging performed on the same day. Volumes were measured in end systole and end diastole and ejection fraction calculated. Right ventricular mass was assessed in end systole. With both methods, the areas of a 2 mm thick slice of the ventricle were manually outlined and multiplied by the slice thickness to obtain slice volume. Slice volumes were multiplied by the number of measured slices to obtain the ventricular volume. Patients-16 patients were studied: three with normal hearts, three after surgical repair of coarctation of the aorta, nine following repair of tetralogy of Fallot, and one with Mustard atrial repair of complete transposition of the great arteries. Results-Correlation between end diastolic volumes measured by both methods was r = 0.95 with limits of agreement ranging from -3.5 to 12.5 ml; correlation for end systolic volumes was r = 0.87 with limits of agreement between -4.0 and 16.4 ml; correlation for end systolic right ventricular mass was r = 0.81 with limits of agreement between - 7.0 and 20.6 g. Interobserver variability ranged from 4.3% (range 0.2% to 9.3%) for end diastolic volume to 7.6% (1.8% to 15.4%) for mass measurements. Conclusions-With transthoracic three dimensional echocardiography, end diastolic right ventricular volumes can be assessed with acceptable accuracy in normal hearts and those with enlarged right ventricles, whereas the current method of three dimensional echocardiography is less good for end systolic volumes and not satisfactory for right ventricular mass measurements.
引用
收藏
页码:127 / 130
页数:4
相关论文
共 13 条
  • [1] STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT
    BLAND, JM
    ALTMAN, DG
    [J]. LANCET, 1986, 1 (8476) : 307 - 310
  • [2] QUANTIFICATION OF RIGHT-VENTRICULAR FUNCTION WITH MAGNETIC-RESONANCE-IMAGING IN CHILDREN WITH NORMAL HEARTS AND WITH CONGENITAL HEART-DISEASE
    HELBING, WA
    REBERGEN, SA
    MALIEPAARD, C
    HANSEN, B
    OTTENKAMP, J
    REIBER, JHC
    DEROOS, A
    [J]. AMERICAN HEART JOURNAL, 1995, 130 (04) : 828 - 837
  • [3] COMPARISON OF ECHOCARDIOGRAPHIC METHODS WITH MAGNETIC-RESONANCE-IMAGING FOR ASSESSMENT OF RIGHT-VENTRICULAR FUNCTION IN CHILDREN
    HELBING, WA
    BOSCH, HG
    MALIEPAARD, C
    REBERGEN, SA
    VANDERGEEST, RJ
    HANSEN, B
    OTTENKAMP, J
    REIBER, JHC
    DEROOS, A
    [J]. AMERICAN JOURNAL OF CARDIOLOGY, 1995, 76 (08) : 589 - 594
  • [4] RIGHT-VENTRICULAR PERFORMANCE AND MASS BY USE OF CINE MRI LATE AFTER ATRIAL REPAIR OF TRANSPOSITION OF THE GREAT-ARTERIES
    LORENZ, CH
    WALKER, ES
    GRAHAM, TP
    POWERS, TA
    [J]. CIRCULATION, 1995, 92 (09) : 233 - 239
  • [5] Martin R W, 1989, J Cardiothorac Anesth, V3, P260, DOI 10.1016/0888-6296(89)90105-1
  • [6] OLDERSHAW P, 1992, BRIT HEART J, V68, P12
  • [7] REDINGTON AN, 1988, BRIT HEART J, V60, P57
  • [8] Rees S, 1988, AM J CARDIOL, V70, P1589
  • [9] COMPARISON OF MAGNETIC-RESONANCE-IMAGING WITH CROSS-SECTIONAL ECHOCARDIOGRAPHY IN THE ASSESSMENT OF LEFT-VENTRICULAR MASS IN CHILDREN WITHOUT HEART-DISEASE AND IN AORTIC ISTHMIC COARCTATION
    VOGEL, M
    STERN, H
    BAUER, R
    BUHLMEYER, K
    [J]. AMERICAN JOURNAL OF CARDIOLOGY, 1992, 69 (09) : 941 - 944
  • [10] 3-DIMENSIONAL ECHOCARDIOGRAPHY CAN SIMULATE INTRAOPERATIVE VISUALIZATION OF CONGENITALLY MALFORMED HEARTS
    VOGEL, M
    HO, SY
    LINCOLN, C
    YACOUB, MH
    ANDERSON, RH
    [J]. ANNALS OF THORACIC SURGERY, 1995, 60 (05) : 1282 - 1288