A sequential injection cold-vapor atomic absorption method for the determination of total mercury

被引:18
作者
Doering, WE
James, RR
Echols, RT [1 ]
机构
[1] Univ Michigan, Dept Civil & Environm Engn, Ann Arbor, MI 48109 USA
[2] Univ Minnesota, Div Sci & Math, Morris, MN 56267 USA
来源
FRESENIUS JOURNAL OF ANALYTICAL CHEMISTRY | 2000年 / 368卷 / 05期
关键词
D O I
10.1007/s002160000512
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A sequential injection (SI) method for the determination of mercury via cold vapor atomic absorption spectrophotometry is presented. The method differs from flow injection (FI) cold vapor methods for the determination of mercury because of the simplicity of the system required for the method: one pump, one valve, a gas-liquid separator, and an atomic absorption spectrophotometer equipped with a quartz cell. Under optimal conditions, the method has the following figures of merit: a linear calibration range of 1.0 to 20 mug L-1; a detection limit of 0.46 mug L-1; and a precision of 0.90% RSD (8 mug L-1). The procedure allows for a sampling rate of one injection per 80 s (excluding sample pretreatment). Results from the determination of mercury in water and fish specimens are also presented. The figures of merit of the method are compared to two other SI methods for the determination of mercury.
引用
收藏
页码:475 / 479
页数:5
相关论文
共 18 条
[1]   A comparison of gas-liquid separators for the determination of mercury by cold-vapour sequential injection atomic absorption spectrometry [J].
Brindle, ID ;
Zheng, SG .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1996, 51 (14) :1777-1780
[2]   Trace determination of mercury: A review [J].
Clevenger, WL ;
Smith, BW ;
Winefordner, JD .
CRITICAL REVIEWS IN ANALYTICAL CHEMISTRY, 1997, 27 (01) :1-26
[3]  
de Mirabó FMB, 1997, ANAL CHIM ACTA, V355, P203
[4]   A CONTINUOUS-FLOW COLD VAPOR PROCEDURE FOR MERCURY DETERMINATION BY ATOMIC-EMISSION USING THE REVERSE FLOW-INJECTION APPROACH [J].
DEANDRADE, JC ;
BUENO, MIMS .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1994, 49 (08) :787-795
[5]   MERCURY DETERMINATION BY COLD VAPOR ATOMIC-ABSORPTION SPECTROMETRY IN SEVERAL BIOLOGICAL INDICATORS FROM LAKE MARACAIBO, VENEZUELA [J].
DEVARGAS, MC ;
ROMERO, RA .
ANALYST, 1992, 117 (03) :645-647
[6]  
Fang Z., 1995, FLOW INJECTION ATOMI
[7]   A time-based flow injection cold vapor atomic absorption spectrometry system with on-line microwave sample pre-treatment for the determination of inorganic and total mercury in urine [J].
Gallignani, M ;
Bahsas, H ;
Brunetto, MR ;
Burguera, M ;
Burguera, JL ;
de Peña, YP .
ANALYTICA CHIMICA ACTA, 1998, 369 (1-2) :57-67
[8]   Determination of mercury in saliva with a flow-injection system [J].
Guo, T ;
Baasner, J ;
Gradl, M ;
Kistner, A .
ANALYTICA CHIMICA ACTA, 1996, 320 (2-3) :171-176
[9]   DETERMINATION OF TOTAL MERCURY IN WATERS AND URINE BY FLOW-INJECTION ATOMIC-ABSORPTION SPECTROMETRY PROCEDURES INVOLVING ONLINE AND OFF-LINE OXIDATION OF ORGANOMERCURY SPECIES [J].
HANNA, CP ;
TYSON, JF ;
MCINTOSH, S .
ANALYTICAL CHEMISTRY, 1993, 65 (05) :653-656
[10]  
Lobring LB., 1991, 2456 EPA