Ultrasound characteristics of subdermally implanted Implanon™ contraceptive rods

被引:27
作者
Lantz, A
Nosher, JL
Pasquale, S
Siegel, RL
机构
[1] Univ Med & Dent New Jersey, Robert Wood Johnson Med Sch, Dept Radiol, New Brunswick, NJ 08903 USA
[2] Univ Med & Dent New Jersey, Robert Wood Johnson Med Sch, Dept Obstet & Gynecol, New Brunswick, NJ 08903 USA
关键词
Implanon (TM) contraceptive rods; subdermally implanted contraceptive device; ultrasound; posterior acoustic shadowing localization;
D O I
10.1016/S0010-7824(97)00159-5
中图分类号
R71 [妇产科学];
学科分类号
100211 ;
摘要
This study was designed to define sonographic characteristics and the optimal sonographic technique for localization of the single rod Implanon(TM) system. Diagnostic ultrasonography was performed in eight women who had the Implanon rod implanted in the medial aspect of the upper arm. All implants were scanned in the longitudinal and transverse direction using 3.5 MHz, 5 MHz, and 7.5 MHz linear array transducers. Scanning was performed both directly against the skin and with 2 and 4 cm Kiteco(TM) ultrasound stand-off pads. Each image obtained was evaluated for detectability of the Implanon rod and the grade of acoustic shadowing produced by the Implanon rod by two independent observers. Discrepancies in evaluation were adjudicated by a third observer. The Implanon rod implant was not directly identified using the transducer and standoff pad combinations. Implanon rods were indirectly identified as a result of the posterior acoustic shadow cast by the Implanon. Best demonstration was achieved with the 5 or 7.5 MHz transducer and a 2 cm stand-off pad in the transverse direction. Appropriate ultrasonographic technique and familiarity with posterior acoustic shadowing patterns generated by the implant provide a noninvasive method for localization of nonpalpable, single rod implants prior to removal. (C) 1997 Elsevier Science Inc. All rights reserved.
引用
收藏
页码:323 / 327
页数:5
相关论文
共 6 条
[1]  
DARNEY PD, 1994, AM J OBSTET GYNECOL, V170, P1536
[2]   CLINICAL-TRIAL WITH 3-KETO-DESOGESTREL SUBDERMAL IMPLANTS [J].
DIAZ, S ;
PAVEZ, M ;
MOOYOUNG, AJ ;
BARDIN, CW ;
CROXATTO, HB .
CONTRACEPTION, 1991, 44 (04) :393-408
[3]   RELEASE KINETICS OF 3-KETO-DESOGESTREL FROM CONTRACEPTIVE IMPLANTS (IMPLANON(R)) IN DOGS - COMPARISON WITH INVITRO DATA [J].
GEELEN, JAA ;
VANDERWARDT, JTH ;
VOORTMAN, G ;
MAASSEN, GCT ;
EENINK, MJD .
CONTRACEPTION, 1993, 47 (02) :215-226
[4]   ULTRASONOGRAPHY - USEFULNESS IN LOCALIZATION OF THE NORPLANT CONTRACEPTIVE IMPLANT SYSTEM [J].
GLAUSER, SJ ;
SCHARLING, ES ;
STOVALL, TG ;
ZAGORIA, RJ .
JOURNAL OF ULTRASOUND IN MEDICINE, 1995, 14 (06) :411-413
[5]   Removal of deeply inserted, nonpalpable levonorgestrel (Norplant(R)) implants [J].
Sarma, SP ;
Wamsher, JG ;
Whitlock, SW .
CONTRACEPTION, 1996, 53 (03) :159-161
[6]   IMAGING OF THE LEVONORGESTREL IMPLANTABLE CONTRACEPTIVE DEVICE [J].
TWICKLER, DM ;
SCHWARZ, BE .
AMERICAN JOURNAL OF OBSTETRICS AND GYNECOLOGY, 1992, 167 (02) :572-573