Measurement of haemoglobin using single drops of skin puncture blood: is precision acceptable?

被引:45
作者
Conway, AM [1 ]
Hinchliffe, RF
Earland, J
Anderson, LM
机构
[1] Univ Sheffield, Dept Paediat, Sheffield Childrens Hosp NHS Trust, Western Bank,Subdept Paediat, Sheffield S10 2TH, S Yorkshire, England
[2] Univ Sheffield, No Gen Hosp, Ctr Human Nutr, Sheffield S5 7AU, S Yorkshire, England
关键词
haemoglobin measurement; skin puncture; blood sampling;
D O I
10.1136/jcp.51.3.248
中图分类号
R36 [病理学];
学科分类号
100104 ;
摘要
The study aimed to investigate local concerns about clinically important discrepancies between repeat HemoCue haemoglobin measurements from single drops of blood. Two biomedical scientists and two health visitors each obtained a series of paired haemoglobin values by fingerprick sampling ham healthy volunteers. Seven of 20 paired values obtained by health visitors and three of 20 obtained by scientists from the first drop of blood forming at the puncture site differed by greater than or equal to 10 g/l; 11 of 20 paired values obtained by health visitors and one of 20 by the scientists from the fourth drop of blood differed by greater than or equal to 10 g/l. After collecting and mixing a number of drops in EDTA tubes before analysis, seven of 40 paired values differed by > 5 g/l, and none by > 10 g/l. Pooling drops of blood before analysis improves precision of HemoCue haemoglobin measurement and allows users to achieve results comparable to those obtained by experienced laboratory staff. Measurement of haemoglobin from single drops of skin puncture blood should be discontinued.
引用
收藏
页码:248 / 250
页数:3
相关论文
共 4 条
[1]  
MILLS A F, 1989, Archives of Disease in Childhood, V64, P1468
[2]   EVALUATION OF PORTABLE HEMOGLOBINOMETER IN GENERAL-PRACTICE [J].
NEVILLE, RG .
BRITISH MEDICAL JOURNAL, 1987, 294 (6582) :1263-1265
[3]  
REEVES JD, 1983, ADV PEDIATR, P281
[4]  
SHARP S, 1995, MDA9521