pH electrode performance under automated management conditions

被引:10
作者
Comer, JEA [1 ]
Hibbert, C [1 ]
机构
[1] Sirius Analyt Instruments Ltd, Forest Row RH18 5DW, E Sussex, England
来源
JOURNAL OF AUTOMATIC CHEMISTRY | 1997年 / 19卷 / 06期
关键词
D O I
10.1155/S1463924697000266
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
PH is frequently measured in laboratories, but to have confidence in the results it is necessary to know that it was measured properly. For an electrode to give accurate results it must be treated well and calibrated correctly. In this Paper, an automated system for PH measurement is described; the system uses the operational pH scale and calibrates using two or three buffer solutions, taking proper account of the effects of temperature on the system. The system can be programmed with standard methods and procedures to ensure that the electrode gives the best possible performance. Calibrations and measurements within the system are reproducible, and the automated system is more robust than the manual pH meter, and requires less operator time.
引用
收藏
页码:213 / 224
页数:12
相关论文
共 8 条
  • [1] BATES RG, 1973, DETERMINATION PH
  • [2] DETERMINATION OF SOIL-PH USING A LABORATORY ROBOT
    BRENES, N
    QUIGLEY, MN
    REID, WS
    [J]. ANALYTICA CHIMICA ACTA, 1995, 310 (02) : 319 - 327
  • [3] COVINGTON AK, 1985, PURE APPL CHEM, V57, P893
  • [4] FRANT MS, 1995, AM LAB, V27, P11
  • [5] Galster H., 1991, PH MEASUREMENT FUNDA
  • [6] Control samples for the determination of pH in low ionic strength waters: Consensus values from interlaboratory tests
    Gardner, MJ
    Ravenscroft, JE
    Ackers, C
    [J]. TALANTA, 1997, 44 (01) : 111 - 117
  • [7] DETERMINATION OF SOIL-PH BY USE OF A ROBOTIC STATION
    TORRES, P
    GARCIAMESA, JA
    DECASTRO, MDL
    VALCARCEL, M
    [J]. FRESENIUS JOURNAL OF ANALYTICAL CHEMISTRY, 1993, 346 (6-9): : 704 - 706
  • [8] Westcott C. C., 1978, pH Measurements