Ultra-performance liquid chromatography coupled to quadrupole-orthogonal time-of-flight mass spectrometry

被引:287
作者
Plumb, R [1 ]
Castro-Perez, J
Granger, J
Beattie, I
Joncour, K
Wright, A
机构
[1] Waters Corp, Milford, MA USA
[2] Waters Corp, MS Technol Ctr, Manchester, Lancs, England
[3] AstraZeneca R&D Charnwood, Phys & Metab Sci, Loughborough, Leics, England
关键词
D O I
10.1002/rcm.1627
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Ultra-performance liquid chromatography (UPLC) utilizes sub-2 mum particles with high linear solvent velocities to effect dramatic increases in resolution, sensitivity and speed of analysis. The reduction in particle size to below 2 mum requires instrumentation that can operate at pressures in the 6000-15 000 psi range. The typical peak widths generated by the UPLC system are in the order of 1-2 s for a 10-min separation. In the present work this technology has been applied to the study of in vivo drug metabolism, in particular the analysis of drug metabolites in bile. The reduction in peak width significantly increases analytical sensitivity by three- to five-fold, and the reduction in peak width, and concomitant increase in peak capacity, significantly reduces spectral overlap resulting in superior spectral quality in both MS and MS/MS modes. The application of UPLC/MS resulted in the detection of additional drug metabolites, superior separation and improved spectral quality. Copyright (C) 2004 John Wiley Sons, Ltd.
引用
收藏
页码:2331 / 2337
页数:7
相关论文
共 11 条
[1]  
Ayrton J, 1998, RAPID COMMUN MASS SP, V12, P217, DOI 10.1002/(SICI)1097-0231(19980314)12:5<217::AID-RCM146>3.0.CO
[2]  
2-I
[3]  
Dear GJ, 1999, RAPID COMMUN MASS SP, V13, P456, DOI 10.1002/(SICI)1097-0231(19990315)13:5<456::AID-RCM508>3.0.CO
[4]  
2-T
[5]   The potential of serially coupled alkyl-bonded silica monolithic columns for high resolution separations of pharmaceutical compounds in biological fluids [J].
Dear, GJ ;
Mallett, DN ;
Higton, DM ;
Roberts, AD ;
Bird, SA ;
Young, H ;
Plumb, RS ;
Ismail, IM .
CHROMATOGRAPHIA, 2002, 55 (3-4) :177-184
[6]  
Eckers C, 1997, RAPID COMMUN MASS SP, V11, P1916, DOI 10.1002/(SICI)1097-0231(199711)11:17<1916::AID-RCM96>3.0.CO
[7]  
2-8
[9]  
LEE MS, 2002, LC MS APPL DRUG DEV
[10]  